Datasheet

www.tektronix.com 7
Timer/Counter/Analyzers—TektronixFCA3000andFCA3100Series
General Specications
Environmenta
l Data
Characteristic
Description
Class MIL-PRF-28800F, Class 3
Operating Temp 0 °C to +50 °C
Storage Temp –40 °C to +71 °C
Humidity
5-95% (10-30 °C)
5-75% (30-40 °C)
5-45% (40-50 °C)
Altitude
Operating: 2,000 m
Storage: 12,000 m
Safety Directive 2006/95/EC, EN61010-1, UL61010-1,
CAN/CSA C22.2 No. 61010-1
EMC EU Direc
tive 2004/108/EC, EN61326-1, EN61326-2-1,
Class A
Power Requirements
Charact
eristic
Description
Basic Version 90 to 265 V
RMS
, 45 to 440 Hz, <40 W
Time-base Options
Characteristic Standard
Medium
Stabil
ity (MS)
High Stability
(HS)
Time-base Type
TCXO OCXO OCXO
Uncertainty Due to
Aging
Per 24h NA
<5×10
–9
*
1
<5×10
–10
*
1
Per month <5×10
–7
<6×10
–8
<1×10
–8
Per year <5×10
–6
<2×10
–7
<5×10
–8
Temperature variation (typ. values)
0-50 °C
<1×10
–5
<5×10
–8
<5×10
–9
20-26 °C
<3×10
–6
<2×10
–8
<1×10
–9
Short-term
Stability: t = 1 s
Not specied
<1×10
–10
<1×10
–11
Root Allan
Variance: t = 10 s
Not specied
<1×10
–10
<1×10
–11
Power-on Stability
NA <1×10
–7
<1×10
–8
Deviation versus
nal value after
24h ON time,
after a warm-up
time of:
30 min 30 min 10 min
Total Uncertainty, for Operating Temperature 20 °C to 26 °C, at 2σ (95%)
Condence Interval
1 year after
calibration
<7×10
–6
<2.4×10
–7
<0.6×10
–7
2 years after
calibration
<1.2×10
–5
<4.6×10
–7
<1.2×10
–7
*
1
After 1 month of continuous operation.
Physical
Dimension
mm
in.
Height 90 3.6
Width 210 8.25
Depth 395 15.6
Weight kg lb.
Net 2.7 5.8
Shipping
3.5
7.5
Ordering Information
Models
Model Description
FCA3000 Timer/Counter/Analyzer 300 MHz / 100 ps
FCA3003 Timer/Counter/Analyzer 3 GHz / 100 ps
FCA3020 Timer/Counter/Analyzer 20 GHz / 100 ps
FCA3100 Timer/Counter/Analyzer 300 MHz / 50 ps
FCA3103 Timer/Counter/Analyzer 3 GHz / 50 ps
FCA3120 Timer/Counter/Analyzer 20 GHz / 50 ps
FCA3000/3100 Series Include: Timer/Counter, line cord, calibration certicate,
Quick Start User Manual, CD-ROM with user manual (English, French, German,
Spanish, Simplied Chinese, Traditional Chinese, Korean, Russian, Japanese),
Programmer’s Guide, Technical Specications, Trial version of TimeView™
Software, and CD-ROM with National Instruments LabVIEW SignalExpress™
Tektronix Edition, Limited Edition Software.
Please specify power plug when ordering.
Instrument Options
Option
Description
MS Medium-stability Oven Time Base
HS High-stability Oven Time Base
RP
Rear-panel Connectors
Power Plug Options
Option
Description
A0 North America
A1 Universal Euro
A2 United Kingdom
A3 Australia
A5
Switzerland
A6 Japan
A10
China
A11 India
E1
UK and Euro (FCA3000 and FCA3100 only)