User Manual

Table Of Contents
Acquire the Sign
al
Standard
CAN 2.0
CAN FD
Bit Rate
500 kbps
FD Bit
Rate
4Mbps
6. Push Standard to select between CAN 2.0
and CAN FD decode/trigger standards.
While CAN FD is
backwards compatible
with CAN 2.0 packets, CAN 2.0 should be
selected for best performance with non CA N
FD bus congu
rations.
FD
Standard
ISO
non-ISO
7. Push Bit R at
e, and turn Multipurp ose a to select from the list of predened bit rates.
Alternatively, you can set the bit rate to a specic value. To do so, select Custom, and then turn M ultipurpose b to set the
bit rate from 10 kbps to 1 Mbps.
When Stand
ard is s et to CAN FD, this will be the S D bit rate.
8. If Standard is set to CAN FD, push FD Bit Rate, and turn Multipurpose a to select from the list of predened bit rates.
Alternat
ively, you can set the bit rate to a specic value. To do so, select Custom, and then turn Multipu rpo se b to set
the bit rate from 500 kbps to 7 Mbps for the MDO3000.
9. If Standa
rd is set to CAN FD, push FD Standard to select between ISO (11898-1:2015) and non-ISO (Bosch:2012)
protocol standards.
76 MDO3000 Series Oscilloscopes User Manual