Datasheet

Level measurement uncertainty Reference level 10 dBm to -15 dBm. Input level ranging from reference level to 40 dB below reference level. Specifications
exclude mismatch error.
18 °C - 28 °C < ±1.2 dBm (< ±0.6 dBm typical)
Over operating range < ±2.0 dBm
Level measurement uncertainty
with TPA-N-PRE preamp attached
Preamp mode set to “Auto”. Reference level 10 dBm set to -40dBm. Input level ranging from reference level to 30 dB below
reference level. Specifications exclude mismatch error.
18 °C - 28 °C < ±1.5 dBm (typical) either preamp state
Over operating range < ±2.3 dBm either preamp state
Frequency measurement accuracy ±(([Reference Frequency Error] x [Marker Frequency]) + (span/750 + 2)) Hz; Reference Frequency Error = 10ppm (10 Hz / MHz)
Maximum operating input level
Average continuous power +20 dBm (0.1 W)
DC maximum before damage ±40 V DC
Maximum power before
damage (CW)
+33 dBm (2 W)
Maximum power before
damage (pulse)
+45 dBm (32 W) (<10 µs pulse width, <1% duty cycle, and reference level of ≥ +10 dBm)
Maximum operating input level
with TPA-N-PRE preamp attached
Average continuous power +20 dBm (0.1 W)
DC maximum before damage ±20 V DC
Maximum power before
damage (CW)
+30 dBm (1 W)
Maximum power before
damage (pulse)
+45 dBm (32 W) (<10 μs pulse width, <1% duty cycle, and reference level of ≥ +10 dBm)
Frequency domain trace types Normal, Average, Max Hold, Min Hold
Detection methods +Peak, -Peak, Average, Sample
Automatic markers One to eleven peaks identified based on user-adjustable threshold and excursion values
Manual markers Two manual markers indicating frequency, amplitude, noise density, and phase noise
Marker readouts Absolute or Delta
FFT windows
FFT window
Factor
Kaiser 2.23
Rectangular 0.89
Hamming 1.30
Hanning 1.44
Blackman-Harris 1.90
Flat-Top 3.77
Datasheet
Spectrum analyzer input
18 www.tektronix.com