User's Manual

AWG510 & AWG520 Service Manual
xvii
Introduction
This manual contains information needed to properly service the AWG510 and
AWG520 Arbitrary Waveform Generators, as well as general information critical
to safe and effective servicing.
To prevent personal injury or damage to the waveform generator, consider the
following before attempting service:
The procedures in this manual should be performed only by a qualified
service person
Read the General Safety Summary and the Service Safety Summary,
beginning on page xi
Read Preparation for Use in section 2, Operating Information
When using this manual for servicing, be sure to follow all warnings, cautions,
and notes.
Performance Check Interval
Generally, the performance check described in section 4,
Performance Verification, should be done every 12 months. In addition, a
performance check is recommended after module replacement.
If the waveform generator does not meet performance criteria, repair is
necessary.
Strategy for Servicing
Throughout this manual, the term, module, refers to any field-replaceable
component, assembly, or part of the waveform generator.
This manual contains all the information needed for periodic maintenance of the
waveform generator. Further, it contains all information for corrective mainte-
nance down to the module level. To isolate a failure to a module, use the
troubleshooting procedures found in section 6, Maintenance. To remove and
replace any failed module, follow the instructions in Removal and Installation
Procedures, also part of section 6. After isolating a faulty module, replace it with
a fully-tested module obtained from the factory. Section 10, Mechanical Parts
List, contains part number and ordering information for all replaceable modules.