151
Appendix A: Specifications
All specifications apply to the TDS1000- and TDS2000-series
oscilloscopes. P2200 probe specifications appear at the end of this
chapter. To verify that the oscilloscope meets specifications, the
oscilloscope must first meet the following conditions:
H The oscilloscope must have been operating continuously for
twenty minutes within the specified operating temperature.
H You must perform the Do Self Cal operation, accessible t hrough
the Utility menu, if the operating temperature changes by more
than 5° C.
H The oscilloscope must be within the factory calibration interval
All specifications are guaranteed unless noted typical.
Oscilloscope Specifications
Acquisition
Acquisition Modes Sample, Peak Detect, and Average
Acquisition Rate,
typical
Up to 180 waveforms per second, per channel (Sample
acquisition mode, no measurements)
Single Sequence Acquisition Mode Acquisition Stops After
g
q
Sample, Peak Detect Single acquisition, all channels
simultaneously
Average N acquisitions, all channels
simultaneously, N is selectable
from 4, 16, 64, and 128