User Guide

Command Descriptions
WFMPre:<wfm>
:NR_Pt? (Query Only)
NOTE. This query is not available when using a TDS210 or TDS220 oscilloscope
(rmware below V 2.00) with a TDS2CMA communications module.
Returns the number of points that are in the transmitted waveform record.
The number of points depends on DATa:STARt, DATa:STOP, and whether
DATa:SOUrce is YT or FFT. NR_Pt is at most 2500 for YT and 1024 for FFT.
NR_Pt is always at least one.
When the DATa:SOUrce is not displayed, the oscilloscope will generate an error
and return event code 2244.
Group
Waveform
Syntax
WFMPre:<wfm>:NR_Pt?
Related Commands
DATa:SOUrce, DATa:STARt
Examples
WFMPre:CH1:NR_Pt
Returns <NR1>, which is the number of data points. If DATa:WIDth is 2, then
there are <NR1>*2 bytes in the curve.
WFMPre:PT_Fmt
The set form of this command sets the format (Y or ENV) of the reference
waveform specied by the DATa:DESTination command.
The query form returns the format of the waveform specied by the DATa:SOUrce
command, if that waveform is on or displayed. If the waveform is not displayed,
the query form of this command generates an error and returns event code 2244.
Group
Waveform
Syntax
WFMPre:PT_Fmt{ENV|Y}
WFMPre:PT_Fmt?
Related Commands
CURVe
2-186 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer