User Guide

Command Descriptions
For the TDS1000
B, TDS2000B, TDS2000C, and TPS2000 Series, <QString>
canalsoreturnthefollowing:
Aforamps
VA for volt-amperes
AA for amperes*amperes
VV for volts*volts
You can also use this command for undenedmathwaveforms.
Examples
MEASUREMENT:IMMED:UNITS? might return "s", indicating that the unit for the
immediate meas urement is seconds.
MEASUrement:IMMed:VALue? (Query Only)
Executes the immediate oscilloscope measurement specied by the
MEASUrement:IMMed:TYPe command. The measurement is taken on the
source specied by the MEASUrement:IMMed:SOUrce command. Immediate
queries and commands are the preferred measurement method for programming.
To check whether the measurement was successful, use the *ESR? query followed
by the EVENT? or ALLEv? queries to read the Standard Event Status Register
(SESR).
NOTE. If the channel specied by MEASUrement:IMMed:SOUrce is not currently
displayed, the oscilloscope generates event 2225 and returns 9.9E37.
If Trigger View is active, Scan mode is in effect, or the display format is set to XY,
this query returns 9.9E37 and generates event 221 (Settings conict)
When math is FFT, tur ned on, and used as a measurement source, attem pting to
query the measurement value returns 9.9e37 and raises error 2225 (no waveform
to measure).
Group
Measurement
Syntax
MEASUrement:IMMed:VALue?
Related Commands
*ESR?, EVENT?,
Returns
<NR3>
2-124 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer