Instruction Manual

ADP30x Probe
ADP300/ADP305 Test Record
Serial Number:
Asset or Tracking Number:
Date:
Technician:
Equipment Used:
MODEL SERIAL NUMBER
CALIBRATION
DUE DATE
OSCILLOSCOPE
DIGITAL MULTIMETER
FUNCTION
GENERATOR
1
1The function generator used in this Performance Verification Procedure is used for making relative
measurements. The output of the generator is measured with a DMM or oscilloscope in this
procedure. Thus, the generator is not required to be calibrated.
Step Description
Intermediate
Data
Test Result
Gain Accuracy
5 Function Generator Output
Voltage
____________V
11
Probe Output Voltage
____________V
13
÷1000 Gain Error (Test limit ±1%)
___________ %
16
Probe Output Voltage
____________V
18
÷100 Gain Error (Test limit ±2%)
___________ %
26 923079-00 Rev A