14
limited to the scale factors that can be obtained without changing the AP033 gain or
attenuation. Thus in Manual gain control mode, only the scale factor of the oscilloscope
will be changed. In this mode, dedicated touch screen controls are displayed for Probe
Attenuation and Probe gain (see Figure 5). The available values that appear in Probe
Attenuation will change depending on the presence of the external attenuator. When the
external ÷10 attenuator is not installed, ÷10 and ÷1 will be displayed. With the external
÷10 attenuator, only ÷100 will be displayed.
When using the AP033 with the ADPPS power supply on oscilloscopes without a ProBus
interface, use Table 2 to select the desired effective volts per division scaling.
Table 2, Recommended settings to obtain desired equivalent Volts/Division
Equivalent
Scale Factor
Oscilloscope
Scale Factor
AP033
Attenuation
AP033
Gain
Common
Mode
Range
Differential
Mode Range
Maximum
Offset
Input
Noise
(nV/√Hz)
100 μV/Div
1 mV/Div
÷1
×
10
±4.2 V
±40 mV
±400 mV
6
200 μV/Div 2 mV/Div ÷1
×
10
±4.2 V ±40 mV ±400 mV 6
500 μV/Div
5 mV/Div
÷1
×
10
±4.2 V
±40 mV
±400 mV
6
1 mV/Div
10 mV/Div
÷1
×
10
±4.2 V
±40 mV
±400 mV
6
2 mV/Div 20 mV/Div ÷1
×
10
±4.2 V ±40 mV ±400 mV 6
5 mV/Div
50 mV/Div
÷1
×10
±4.2 V
±40 mV
±400 mV
6
10 mV/Div
100 mV/Div
÷1
×
10
±4.2 V
±40 mV
±400 mV
6
20 mV/Div 20 mV/Div ÷10
×
10
±42 V ±400 mV ±4 V 60
50 mV/Div
50 mV/Div
÷10
×10
±42 V
±400 mV
±4 V
60
100 mV/Div
100 mV/Div
÷10
×10
±42 V
±400 mV
±4 V
60
200 mV/Div
20 mV/Div
÷10
×
1
±42 V
±4 V
±4 V
115
500 mV/Div 50 mV/Div ÷10
×
1
±42 V ±4 V ±4 V 115
1 V/Div
100 mV/Div
÷10
×1
±42 V
±4 V
±4 V
115
2 V/Div
20 mV/Div
÷100
1
×
1
±42 V
±40 V
±40 V
1150
5 V/Div 50 mV/Div ÷100
1
×
1
±42 V ±40 V ±40 V 1150
10 V/Div
100 mV/Div
÷100
1
×1
±42 V
±40 V
±40 V
1150
1. ÷100 attenuation is obtained using the external ÷10 attenuator. The probe is automatically set to ÷10
internally when the external attenuator is installed. The internal ÷1 attenuator cannot be selected when
the external ÷10 attenuator is installed.