Instructions

Performance Verification
CP030OM-E Rev B ISSUED: September 2005 5-1
5Performance Verification
This procedure can be used to verify the warranted characteris-
tics of the CP030 Current Probe.
The recommended calibration interval for the model CP030 Cur-
rent Probe is one year. The complete performance verification
procedure should be performed as the first step of annual calibra-
tion. Test results can be recorded on a photocopy of the Test
Record provided at the end of the manual.
Performance verification can be completed without removing the
probe covers or exposing the user to hazardous voltages. Adjust-
ment should only be attempted if a parameter measured in the
Performance Verification Procedure is outside the specification
limits.
Adjustment should only be performed by qualified personnel.
TEST EQUIPMENT REQUIRED
The following table lists the test equipment and accessories (or
their equivalents) which are required for performance verification
of the CP030 Current Probe.
Because the input and output connector types may vary on differ-
ent brands and models of test instruments, additional adapters or
cables may be required.
Table 5-1. List of Required Equipment
Description Minimum Requirements Test Equipment Examples
Wide Band Oscilloscope Minimum 200 MHz bandwidth
ProBus interface equipped
Software version 4.3.0.0 or higher
LeCroy WaveRunner 6030A
Digital Multimeter
(2 required)
DC: 0.1% Accuracy
5½ digit resolution
Agilent Technologies 34401A or
Fluke 8842A-09
Function Generator 50 Hz sine wave output. 3 Vrms into
50
Agilent Technologies 33120A or
Stanford Research Model DS340
Calibration Fixture, 100 Turn
Loop
100 Turn loop in series with 0.5
±
0.1% resistor with sense termi-
nals.
LeCroy CP030-CF02
Calibration Fixture ProBus Extension Cable LeCroy PROBUS-CF01