Instructions
Performance Verification
CP030OM-E Rev B ISSUED: September 2005 5-3
PERFORMANCE VERIFICATION PROCEDURE
The warranted characteristics of the CP030 Current Probe are
valid at any temperature within the Environmental Characteristics
listed in the Specifications. However, some of the other test
equipment used to verify the performance may have environmen-
tal limitations required to meet the accuracy needed for the pro-
cedure. Make sure that the ambient conditions meet the
requirements of all the test instruments used in his procedure.
Note
The correct operation of the CP030 controls requires
software version 4.3.0.0 or higher. The software version
in the test oscilloscope can be verified by selecting
Utilities, Utilities Setup, then the Status tab. Contact
your local LeCroy representative if the software in your
oscilloscope requires updating.
CHECK LF ACCURACY
1. Set the Function generator to 50 Hz sine wave. Output volt-
age at 3 Vrms with 50Ω output.
Figure 5-1. LF Accuracy Test Set Up