Instructions

CP030 Current Probe
1-2 ISSUED: September 2005 CP030-OM-E Rev B
USAGE
The probe is intended to be used only with instruments that are
connected to earth ground through the input connector. The
probe is not intended to be used in wet or explosive
atmospheres.
The use of the probe and/or the instrument it is connected to in a
manner other than that specified may impair the safety
mechanisms.
In order to assure safe operation and to obtain maximum
performance from the unit, observe the following:
1. To avoid short circuits and accidents that could result in injury
or death, use the CP030 current probe only with conductors
carrying 300 V or less.
2. In order to prevent short circuits and electric shock when con-
ductors being measured carry less than 300 V but more than
the safe voltage level (SELV-E) make sure that the conduc-
tors to be measured are insulated with material conforming
to:
Overvoltage Category I
Basic insulation requirement for working voltage of 300 V.
Pollution degree 2
3. Never install or remove the probe on bare conductors which
are energized. The transformer core and shield are grounded
but not insulated and may contact the conductor when the
locking lever is open.
4. Be careful not to damage the insulation surface when making
measurements.
5. Do not use the probe if any part is damaged. All maintenance
should be referred to qualified service personnel.
CAUTION
WARNING