Instructions

Specifications
CP150-OM-E Rev C 7-1
7Specifications
NOMINAL CHARACTERISTICS
Nominal characteristics describe parameters and attributes that
are guaranteed by design, but do not have associated tolerances.
Maximum Continuous Input 150 Amp rms
Current Refer to Figure 7-1, “Input Cur-
rent vs. Frequency.”
Maximum Non-Continuous 300 Amp peak,
Peak Current 500 Amp at pulse width of
< 30 µsec.
Insertion Impedance Refer to Figure 7-2.
Intended Output Load 1 M
Maximum Permitted 600 V Cat II, 300 V Cat III
Circuit Voltage (Insulated conductor)
WARRANTED CHARACTERISTICS
Warranted characteristics describe parameters that have
guaranteed performance. Unless otherwise noted, tests are
provided in the Performance Verification Procedure for all
warranted specifications.
Guaranteed at 23 °C ±3 °C (73 °F ±5 °F) after power has been
applied for 30 minutes.
Low Frequency Accuracy ±1.0% of reading ±100 mA.
TYPICAL CHARACTERISTICS
Typical characteristics describe parameters which do not have
guaranteed performance, however are representative of the
average performance from a sample of several probes. Tests for
typical characteristics are not provided in the Performance
Verification procedure.
Sensitivity* 200 mA/div* to 100 A/div
Output voltage 0.01 V/A
* Lower sensitivity may change with different oscilloscope models.