Operator's Manual
Back Input/Output Panel
A. Aux Out connector sends device trigger enabled, trigger out, or pass/fail output to another device.
B. Ref In/Out connector allows you to input an external Reference Clock, or to output a Reference Clock
to another instrument.
C. USBTMC Port enables remote control of the oscilloscope.
D. AC Power Inlet connects the AC line cord.
See the general set up instructions for more information about configuring connections to other devices.
Analog Inputs
A series of BNC connectors arranged on the front and back of the oscilloscope are used to input analog
signal on Channels 1-4 and AUX In, or an external trigger pulse on EXT.
HDO connectors use the ProBus interface and are compatible with any Teledyne LeCroy ProBus type
probes rated for the oscilloscope's bandwidth.
The ProBus interface contains a 6-pin power and communication connection and a BNC signal connection
to the probe. It offers both 50 /1 M input impedance and provides probe power and control for a wide
range of probes such as high impedance passive probes, high impedance active probes, current probes,
high voltage probes, and differential probes. ProBus also includes sense rings for detecting passive
probes. The ProBus interface may also have a BNC-terminated cable connected directly to it.
The interfaces power probes and completely integrate the probe with the oscilloscope channel. Upon
connection, the probe type is recognized and some setup information, such as input coupling and
attenuation, is performed automatically. This information is displayed on the Channel Probe Dialog.
System (probe plus oscilloscope) gain settings are automatically calculated and displayed based on the
probe attenuation.
Probes
HDO6000 oscilloscopes are compatible with the included passive probes and all Teledyne LeCroy ProBus
active probes that are rated for the oscilloscopes bandwidth. Probe specifications and documentation
are available at teledynelecroy.com/hdo6000.
The passive probes supplied with your oscilloscope are matched to the input impedance of the
instrument, but may need further compensation; refer to the probe manual for the procedure. If using
other passive probes with your oscilloscope, be sure to perform a low frequency calibration using the Cal
signal available from the HDO's front panel before using them to measure signal.Follow the directions in
the probe instruction manual to compensate the low and/or high frequency response of the probes.