Operators Manual Owner manual

Operator's Manual
Sampling Modes
Real Time Sampling Mode
Real Time sampling mode is a series of digitized voltage values sampled on the input signal at a uniform
rate. These samples are displayed as a series of measured data values associated with a single trigger
event. By default, the waveform is horizontally positioned so that the trigger event is time zero on the grid.
The relationship between sample rate, memory, and time can be expressed as:
Capture Interval = 1/Sample Rate X Memory
Capture Interval/10 = Time Per Division
In Real Time sampling mode, the acquisition can be displayed for a specific period of time (or number of
samples) either before or after the trigger event occurs, known as trigger delay. This allows you to isolate
and display a time/event of interest that occurs before or after the trigger event.
l Pre-trigger delay displays the time prior to the trigger event. This can be set from a time well before
the trigger event to the moment the event occurs, up to the oscilloscope's maximum sample record
length. How much actual time this represents depends on your timebase setting. When set to the
maximum allowed pre-trigger delay, the trigger position (and zero point) is off the grid (indicated by
the trigger delay arrow at the lower right corner), and everything you see represents pre-trigger
time.
l Post-trigger delay displays time following the trigger event. Post-trigger delay can cover a much
greater lapse of time than pre-trigger delay, up to the equivalent of 10,000 time divisions after the
trigger event occurred. When set to the maximum allowed post-trigger delay, the trigger point may
actually be off the grid far to the left of the time displayed.
Usually, on fast timebase settings, the maximum sample rate is used when in Real Time mode. For
slower timebase settings, the sample rate is decreased so that the maximum number of data samples is
maintained over time.
Roll Mode
Roll mode displays, in real time, incoming points in single-shot acquisitions that have a sufficiently low
data rate. This mode can be invoked for slow acquisitions where the time per division is 100 ms/div or
slower. Roll mode samples at 2.5 MS/s.
The oscilloscope appears to "roll" the incoming data continuously across the screen until a trigger event
is detected and the acquisition is complete. The parameters or math functions connected to each
channel are updated every time the roll mode buffer is updated, as if new data is available. This resets
statistics on every step of Roll mode that is valid because of new data.
NOTE: If the processing time is greater than the acquire time, the data in memory is overwritten. In this
case, the instrument issues the warning, "Channel data is not continuous in ROLL mode!!!" and rolling
starts again.
RIS Sampling Mode
RIS (Random Interleaved Sampling) is an acquisition technique that allows effective sampling rates
higher than the maximum single-shot sampling rate. It is used on repetitive waveforms with a stable
trigger. The maximum effective RIS sampling rate is achieved by making multiple single-shot
acquisitions at maximum real-time sample rate. The bins thus acquired are positioned approximately 8
922499 Rev B
35