Table of measurements open beneath grid. Far left cell opens the Measure dialog.
To quickly access the Measure Setup dialog if it is closed, touch the far left cell of the readout table
labeled Measure.
View Statistics
You can add the statistical measures value(last), mean, min., max., sdev, and num(ber of measurements
computed) to the measurement parameter readout table.
To turn on statistics, access the Measure dialog and check Statistics On.Clear the checkbox to remove
statistics from the readout. You can also choose Measure > Statistics from the menu bar.
The num statistic is the number of measurements computed. For any parameter that computes on an
entire waveform (like amplitude, mean, minimum, maximum, etc.) the value displayed represents the
number of sweeps.
For any parameter that computes on every event, the value displayed is equal to the number of events per
acquired waveform. If x waveforms were acquired, the value represents x times the number of cycles per
waveform. The value(last) statistic is equal to the measurement of the last cycle on the last acquisition.
To reset the statistics counter, touch Clear Sweeps on the display or Front Panel.
View Histicon
Histicons are miniature histograms of measurement parameters that appear on the measurement table.
These thumbnail histograms let you see at a glance the statistical distribution of each parameter.
1. Choose Measure > Measure Setup from the menu bar to access the Measure dialog.
2. Select Show Table.
3. Check On to enable the parameters you wish to display.
4. Select Statistics Histicons.
NOTE: You can quickly display a full histogram by touching the histicon you want to enlarge. The enlarged
histogram appears superimposed over its source trace.
922499 Rev B