Datasheet

3
WORLD’S ONLY 2 KV SAFETY RATED MODEL
Solar PV Inverters
The HVD3206A probe is rated to 1500 V
dc
(2000 V DC + peak AC). This makes the probe ideal for testing single-phase
or three-phase inverters or newer 1500 V
dc
rated string-inverters (per IEC/EN 61010-031:2015). The HVD3206A has
the same excellent 1% gain accuracy, industry-leading CMRR, industry-best offset range, standard 2 meter cable (6
meter cable model available), and a guaranteed 2000 V
pk
differential voltage range.
Only 1500 V
dc
safety rated probe per
IEC/EN 61010-031:2015
Guaranteed 2000 V
pk
differential
voltage range
Industry’s best offset capability
(1500 V) when used with HDO Series
oscilloscopes
Bandwidth rating up to 120 MHz
6 meter cable model available
(HVD3206A-6M)
DC-DC Converters
High-power DC-DC converters can operate at substantial voltages, 500 V
dc
or higher. The HVD310xA models provide up to
1000 V
dc
common-mode (HVD3206A models provide up to 1500 V
dc
) and high precision (1% gain accuracy) DC voltage
measurements. Automatic switchable attenuation keeps the probe in the optimum measurement range. Multiple probes
can be used to understand complex device switching performance.
Common mode range up to ±2000 V
(DC + peak AC) with HVD3206A
High precision - 1% accuracy
Automatic gain switching for optimum
performance
*Based on 30% overshoot condition
Solar String DC
Bus Voltage
Sinusoidal AC Output
V
dc
V
pk
(Rated) V
pk
(with Overshoot)*
Line-Line
Line-
Neutral
Line-Line
Line-
Neutral
1000 849 490 1104 637
1500 849 490 1104 637