Datasheet

6
SPECIFICATIONS
HVD3102A HVD3106A HVD3106A-6M
Bandwidth 25 MHz 120 MHz 80 MHz
Rise Time (10-90) 14 ns 2.9 ns 4.4 ns
Differential Voltage
Range (High Attenuation)
1500 V (DC + peak AC) from 7 to 500 V/
div with up to 1500 V offset.
1750 V maximum typical measurable
differential voltage before saturation.
1500 V (DC + peak AC) from 7 to 500 V/div with up to 1500 V offset.
2000 V maximum typical measurable differential voltage before saturation
Differential Voltage
Range (Low Attenuation)
27.6 V (DC + peak AC) from 100 mV/div to 6.9 V/div with up to 150 V offset
Common Mode Voltage
Range
±1500 V (DC + peak AC), 1000 V
rms
Maximum Input Voltage
to Earth
±1500 Vpk (nominal, either input to ground)
Max Safe Input Voltage 1000 V
rms
CAT III
Sensitivity 100 mV/div to
6.9V/div (100X)
7V/div to 500V/div (1000X)
100 mV/div to
6.9V/div (50X)
7V/div to 500V/div (500X)
100 mV/div to 6.9V/div (50X)
7V/div to 500V/div (500X)
Gain Accuracy 1% (LF, guaranteed)
Slew Rate 100 V/ns (maximum) 400 V/ns (maximum) 270 V/ns (maximum)
Attenuation 100x / 1000x 50x / 500x 50x / 500x
Input Impedance 10 MΩ || 2.5 pF (between inputs), 5 MΩ || 5.0 pF (either input to ground)
Input Coupling DC only
Output Coupling AC or DC coupling
Output Termination 1 MΩ
Interface ProBus
Input Lead Length 40 cm input lead length
Cable Length (input lead
to oscilloscope
connection)
2.25 m 6.8 m
Noise and Rejection
CMRR (Typical) DC - 60 Hz: 85 dB
1 MHz: 65 dB
5 MHz: 40 dB
20 MHz: 30 dB
DC - 60 Hz: 85 dB
1 MHz: 65 dB
5 MHz: 40 dB
20 MHz: 30 dB
100 MHz: 30 dB
DC - 60 Hz: 85 dB
1 MHz: 65 dB
5 MHz: 40 dB
20 MHz: 30 dB
80 MHz: 30 dB
Noise (Probe) 100X: <15 mV
rms
1000X: <85 mV
rms
(referred to input)
50X: <30 mV
rms
500X: <150 mV
rms
(referred to input)
50X: <30 mV
rms
500X: <150 mV
rms
(referred to input)
Environmental
Temperature (Operating) 0°C to 50°C
Temperature
(Non-Operating)
-40°C to 70°C
Humidity (Operating) 5% to 80% RH (Non-Condensing) up to 30°C, decreasing linearly to 45% RH at 50°C
Humidity (Non-Operating) 5% to 95% RH (Non-Condensing), 75% RH above 30°C, 45% RH above 40°C
Altitude (Operating) 3000 m maximum
When used with clip accessories, 2000 m maximum
Altitude (Non-Operating) 10,000 m
Pollution Degree 2, Indoor use only
Certications
CE (LVD Directive
2006/95/EC)
IEC/EN 61010-031:2015
CE (EMC Directive
2004/108/EC)
IEC/EN 61326-1:2013
UL Listed UL 61010-031 (Second Edition)
cUL Listed CAN/CSA-C22.2 No. 61010-031-15