Datasheet

7
HVD3206A HVD3206A-6M HVD3605A
Bandwidth 120 MHz 80 MHz 100 MHz
Rise Time (10-90) 2.9 ns 4.4 ns 4.3 ns
Differential Voltage
Range (High Attenuation)
2000 V (DC + peak AC) from 7 to 500
V/div
with up to 1500V offset.
2000 V (DC + peak AC) from 7 to 500
V/div
with up to 1500V offset.
7000 V (DC + peak AC) from 28V/div to
2000 V/div
with up to 6000V offset.
(7600 V maximum measurable
differential voltage before saturation)
Differential Voltage
Range (Low Attenuation)
27.6 V (DC + peak AC) from 100 mV/div
to 6.9 V/div
with up to 150V offset.
27.6 V (DC + peak AC) from 100 mV/div
to 6.9 V/div
with up to 150V offset.
700 V (DC + peak AC) from 300mV/div
to 27.5 V/div
with up to 600V offset.
Common Mode Voltage
Range
±2000 V (DC + peak AC) ±2000 V (DC + peak AC) ±7600V (DC + peak AC), 6000 Vrms
Maximum Input Voltage
to Earth
±2000 Vpk
(nominal, either input to ground)
±2000 Vpk
(nominal, either input to ground)
±7600 Vpk
(nominal, either input to ground)
Max Safe Input Voltage 2000 V (DC + peak AC) CAT I *
1500 Vdc CAT III
1000 Vrms CAT III
2000 V (DC + peak AC) CAT I *
1500 Vdc CAT III
1000 Vrms CAT III
8485 V (DC + peak AC) CAT I *
6000 Vrms CAT I *
1500 Vdc CAT III
1000 Vrms CAT III
Sensitivity 100 mV/div to 6.9V/div (50X)
7V/div to 500V/div (500X)
100 mV/div to 6.9V/div (50X)
7V/div to 500V/div (500X)
300 mV/div to 27.5 V/div (200X)
28 V/div to 2000 V/div (2000X)
Gain Accuracy 1% (LF, guaranteed)
Slew Rate 400 V/ns (maximum) 270 V/ns (maximum) 1000 V/ns (typical)
Attenuation 50x / 500x 50x / 500x 200x / 2000x
Input Impedance 10 MΩ || 2.5 pF (between inputs)
5 MΩ || 5.0 pF (either input to ground)
10 MΩ || 2.5 pF (between inputs)
5 MΩ || 5.0 pF (either input to ground)
48 MΩ || 2.5 pF (between inputs)
24 MΩ || 5.0 pF (either input to ground)
Input Coupling DC only
Output Coupling AC or DC coupling
Output Termination 1 MΩ
Interface ProBus
Input Lead Length 40 cm input lead length
Cable Length (input lead
to oscilloscope
connection)
2.25 m 6.8 m 6.8 m
Noise and Rejection
CMRR (Typical) DC - 60 Hz: 85 dB
1 MHz: 65 dB
5 MHz: 40 dB
20 MHz: 30 dB
100 MHz: 30 dB
DC - 60 Hz: 85 dB
1 MHz: 65 dB
5 MHz: 40 dB
20 MHz: 30 dB
100 MHz: 30 dB
DC - 60 Hz: 85 dB
10 kHz: 70 dB
1 MHz: 64 dB (200x)
1 MHz: 50 dB (2000x)
10 MHz: 40 dB (200x)
10 MHz: 30 dB (2000x)
100 MHz: 30 dB
Noise (Probe) 50X: <30 mV
rms
500X: <150 mV
rms
(referred to input)
50X: <30 mV
rms
500X: <150 mV
rms
(referred to input)
200x: <65 mV
rms
2000x: <320 mVrms
(typical, referred to input)
Environmental
Temperature (Operating) 0°C to 50°C
Temperature
(Non-Operating)
-40°C to 70°C
Humidity (Operating) 5% to 80% RH (Non-Condensing) up to 30°C, decreasing linearly to 45% RH at 50°C
Humidity (Non-Operating) 5% to 95% RH (Non-Condensing), 75% RH above 30°C, 45% RH above 40°C
Altitude (Operating) 3000 m maximum
When used with clip accessories, 2000 m maximum
Altitude (Non-Operating) 10,000 m
Pollution Degree 2, Indoor use only
Certications
CE (LVD Directive
2006/95/EC)
IEC/EN 61010-031:2015
CE (EMC Directive
2004/108/EC)
IEC/EN 61326-1:2013
UL Listed UL 61010-031 (Second Edition)
cUL Listed CAN/CSA-C22.2 No. 61010-031-15
* CAT I per IEC/EN 61010-031/A1:2008. No Rated Measurement Category per IEC/EN 61010-031:2015.