Data Sheet
© 2019 Teledyne LeCroy, Inc. All rights reserved. Specications, prices, availability, and delivery subject to change without notice.
Product or brand names are trademarks or requested trademarks of their respective holders.
hvfo-probes-ds-05feb19
Local sales oces are located throughout the world.
Visit our website to nd the most convenient location.
1-800-5-LeCroy
teledynelecroy.com
Customer Service
Teledyne LeCroy oscilloscopes and probes are designed, built, and tested to ensure high reliability. In the unlikely event you experience difculties,
our digital oscilloscopes are fully warranted for three years and our probes are warranted for one year. This warranty includes:
•
No charge for return shipping
•
Long-term 7-year support
•
Upgrade to latest software at no charge
SPECIFICATIONS & ORDERING INFORMATION
Product Description Product Code
High Voltage Fiber Optically-isolated Probe Models and Accessories
High Voltage Fiber Optic Probe, 60 MHz Bandwidth. Includes soft-carrying case, Qty. 1 Amplier/Modulating Transmitter, Qty. 1
Demodulating Receiver, Qty. 1 1m Fiber Optic Cable, Qty. 1 USB Charging Cable, Qty. 1 Micro-gripper set.
Attenuating Tips must be ordered separately.
HVFO103
HVFO10X +/-1V (1x Attenuation) Universal Tip Accessory HVFO100-1X-TIP-U
HVFO10X +/-5V (5x Attenuation) Universal Tip Accessory HVFO100-5X-TIP-U
HVFO10X +/-10V (10x Attenuation) Universal Tip Accessory HVFO100-10X-TIP-U
HVFO10X +/-20V (20x Attenuation) Universal Tip Accessory HVFO100-20X-TIP-U
HVFO10X +/-40V (40x Attenuation) Universal Tip Accessory HVFO100-40X-TIP-U
Spare Amplier/Modulating Transmitter (permits constant operation - charge one while another is being used) HVFO103-XMITTER
1m Spare Fiber Optic Cable Accessory HVFO-1M-FIBER
2m Fiber Optic Cable Accessory HVFO-2M-FIBER
6m Fiber Optic Cable Accessory HVFO-6M-FIBER
NIST Traceable Calibration Certicate HVFO103-CCNIST
Electrical
Bandwidth 60 MHz (typical, with tip attached)
Rise Time (10-90%) 7.5 ns (typical)
Input Dynamic Range ±1V, ±5V, ±10V, ±20V, ±40V (DC+peak AC) respectively with 1X, 5X, 10X, 20X or 40X attenuating tips.
All tips are purchased as accessories (none are included with HVFO103 probe).
Maximum Non-destruct Voltage 5 times the operating voltage rating (tip dependent)
Common Mode Voltage Range ±35 kV (DC+Peak AC) (not for hand-held use, with adequate spacing between probe components and earth ground)
Maximum Input Voltage to Earth ±35 kV (DC+Peak AC) (not for hand-held use, with adequate spacing between probe components and earth ground)
Maximum Safe Input Voltage For hand-held use, 30 Vrms / 60 Vdc per IEC/EN 61010-031:2015
Offset Offset capability determined by oscilloscope offset available in a given gain (V/div) setting after accounting for total
probe attenuation (total probe attenuation is twice the tip attenuation).
Sensitivity 50 mV/div to 1 V/div (1X tip), 250 mV/div to 5 V/div (5X tip), 500 mV/div to 10 V/div (10X tip), 1 V/div to 20 V/div (20X tip),
2 V/div to 40 V/div (40X tip)
Gain Accuracy 2% (LF, guaranteed)
Input Impedance 1 MΩ || 34 pF (1X tip); 5 MΩ || 26 pF (5X tip); 8 MΩ || 23 pF (10X tip); 10 MΩ || 22 pF (20X tip); 10 MΩ || 22 pF (40X tip)
Input/Output Coupling DC only
Interface ProBus
Cable Length 1.25 m (4.1 feet) from input lead to oscilloscope connection (using included 1 meter ber optic cable)
Battery 6 hour battery life (typical). 1.5 hour re-charge time (typical, with user-supplied dedicated USB charger).
3 hour re-charge time (typical) using supplied USB charging cable connected to oscilloscope USB port
Noise, Rejection, and Electromagnetic Compatibility (EMC)
CMRR (typical) 140 dB (100 Hz), 120 dB (to 1 MHz), 85 dB (to 10 MHz),
60 dB (to 60 MHz)
Noise (Probe only) 7 mVrms (1X tip), 35 mVrms (5X tip), 70 mVrms (10X tip), 140 mVrms (20X tip), 280 mVrms (40X tip)
Electrostatic Discharge (ESD)
Immunity
8 kV contact discharge and 10 kV air discharge per IEC61000-4-2, criteria A
Radiated RF Electromagnetic
Field Immunity
25 V/m (80 MHz to 2.7 GHz) per IEC61000-4-3, criteria A
Immunity to Conducted
Disturbance Induced by
RF Fields
10 V/m (150 kHz to 80 MHz) per IEC61000-4-6, criteria A
Environmental
Temperature 0°C to 50°C (operating), -20°C to 70°C (non-operating)
Humidity 5% to 80% RH (non-condensing) up to 30°C, decreasing linearly to 45% RH at 50°C (operating)
5% to 95% RH (non-condensing), 80% RH above 30°C, 45% RH above 50°C (non-operating)
Altitude Up to 3000 m (operating), 10,000 m (non-operating)
Pollution Degree 2, Indoor Use Only
Certications
CE Declaration of Conformity Low Voltage Directive 2014/35/EU (IEC/EN 61010-031:2015
EMC Directive 2014/30/EU (IEC/EN 61326-1:2013)
RoHS2 Directive 2011/65/EU
Ordering Information