PP017/PP018 Passive Probe Instruction Manual
700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 www.teledynelecroy.com PP017/PP018 Passive Probe Instruction Manual © 2012 Teledyne LeCroy Corporation. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales and distribution purposes is strictly prohibited. However, clients are encouraged to distribute and duplicate Teledyne LeCroy documentation for their own internal educational purposes.
Instruction Manual Warranty NOTE: THE WARRANTY BELOW REPLACES ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. TELEDYNE LECROY SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT OR OTHERWISE. THE CUSTOMER IS RESPONSIBLE FOR THE TRANSPORTATION AND INSURANCE CHARGES FOR THE RETURN OF PRODUCTS TO THE SERVICE FACILITY.
PP017/PP018 Passive Probe Terms and Symbols The following terms appear in this manual: Warnings identify conditions or practices that could result in injury or loss of life.
Instruction Manual Specifications Model Input Imped. Attn. B/W Rise Time Length Compensation Range R (MΩ) C (pF) (MHz) (ns) (m) (pF) PP017 10x 10 12 250 1.4 1.3 10 - 30 PP018 10x 10 10 500 0.7 1.3 6 - 22 Attenuation Ratio................................... 10x (10:1) 10:1 ±0.5% Max. Input Voltage................................. 600 V CAT I or 300 V CAT II (DC + peak AC) ...................................................................
PP017/PP018 Passive Probe EC Declaration of Conformity Compliance was demonstrated to the specification as listed in the Official Journal of the European Communities: Low Voltage Directive 2006/95/EC. EN 61010-031/A1:2008 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 031: Safety requirements for handheld probe assemblies for electrical measurement and test.
Instruction Manual Do not use in wet or explosive atmospheres. Avoid physical injury. The probe tip is extremely sharp. Use care when handling to prevent injury, including accidental skin puncture. For indoor use only. Cleaning Use the following guidelines: Clean only the exterior of the probe, cables, and accessories. Use a soft cotton cloth moistened with a mild detergent and water solution. Do not allow any portion of the probe to be submerged at any time.
PP017/PP018 Passive Probe Probe Compensation Proper compensation of the probe is required to assure amplitude accuracy of the waveform being measured by matching the probe to the oscilloscope’s input capacitance. Compensation should be adjusted whenever the probe is connected to or transferred between oscilloscopes. Low Frequency (LF) 1. Apply a 1 kHz square wave to the probe or connect to the oscilloscope’s calibrator output. 2.
Instruction Manual High Frequency (HF) High frequency should rarely require adjustment. However, if adjustment is required: 1. Apply a 1 MHz square wave to the probe (< 0.7 ns rise-time). 2. Adjust the right trimmer (HF) located on the PP017probe body: or both trimmers on the PP018 BNC Box (HF2 or HF1): until you achieve a flat-topped square wave per the center illustration in the figure below.
PP017/PP018 Passive Probe PP017 Input Impedance Profile PP018 Input Impedance Profile 8 920876-00 Rev A
Instruction Manual Voltage vs.
PP017/PP018 Passive Probe 10 920876-00 Rev A
Instruction Manual Replaceable Parts Item 1 2 3 4 5 6 7 8 9 Description Sprung Hook, 5 mm, Black GND Lead w/Alligator Clip BNC Adapter, 5 mm Replacement Tip, Black Deluxe Trimmer Tool IC Tip Insulator, 5 mm, Black Tip Insulator, 5 mm, Black Identifier Rings Probe Tip GND Quantity 1 1 1 1 1 1 1 1 1 Standard Accessory Replacement Kit part number is PKIT4-5MM-101.
PP017/PP018 Passive Probe Contact Teledyne LeCroy Contact your local Teledyne LeCroy office for sales and support. Teledyne LeCroy Offices World Wide United States and Canada - World Wide Corporate Office Teledyne LeCroy Corporation 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Ph: 800-553-2769 (Sales 1, Applications 2, Service 3) or 845-425-2000 Fax: 845-578-5985 contact.corp@teledynelecroy.com www.teledynelecroy.
Instruction Manual China Teledyne LeCroy Beijing Rm. 2001 Unit A, Horizon Plaza No. 6, Zhichun Rd., Haidian Dist. Beijing 100088, China Ph: ++86 10 8280 0318/0319/0320 Fax: +86 10 8280 0316 www.teledynelecroy.com.cn Service: Rm. 2002 Ph: ++86 10 8280 0245 Taiwan LeColn Technology Co Ltd. Far East Century Park, C3, 9F No. 2, Chien-8th Road, Chung-Ho Dist., New Taipei City, Taiwan Ph: ++ 886 2 8226 1366 Fax: ++ 886 2 8226 1368 sales_twn@teledynelecroy.com.tw www.teledynelecroy.com.