PP019/PP020 Passive Probe Instruction Manual
700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com PP019/PP020 Passive Probe Instruction Manual © 2014 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales and distribution purposes is strictly prohibited. However, clients are encouraged to distribute and duplicate Teledyne LeCroy documentation for their own internal educational purposes.
Instruction Manual Warranty NOTE: THE WARRANTY BELOW REPLACES ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. TELEDYNE LECROY SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT OR OTHERWISE. THE CUSTOMER IS RESPONSIBLE FOR THE TRANSPORTATION AND INSURANCE CHARGES FOR THE RETURN OF PRODUCTS TO THE SERVICE FACILITY.
PP019/PP020 Passive Probe Terms and Symbols The following terms appear in this manual: CAUTION of potential damage to equipment, or WARNING of potential for bodily injury. Attend to the accompanying information to protect against personal injury or damage. Do not proceed until conditions are fully understood and met. The following symbols may appear on the product: DANGER High Voltage, risk of electric shock or burn.
Instruction Manual Specifications Model Input Imped. Attn. B/W Rise Time Length Compensation Range R (MΩ) C (pF) (MHz) (ns) (m) (pF) PP019 10x 10 12 250 1.4 1.2 10 - 30 PP020 10x 10 11 500 0.7 1.2 8 - 20 Attenuation Ratio .....................10:1 ± 1% Max. Input Voltage ...................Measurement Category I: 500 Vrms CAT I 1500 V transient overvoltage (see voltage derating curve on page 10) Measurement Category II: 400 Vrms CAT II Pollution Degree ......................
PP019/PP020 Passive Probe EC Declaration of Conformity The product is marked with this symbol to indicate that it conforms to the applicable European Union requirements per Low Voltage Directive (LVD) 2006/95/EC.
Instruction Manual Do not overload. Do not apply any potential to the probe leads that exceeds the maximum rating of the probe. Always comply with the Voltage vs. Frequency Derating Curve. Observe all terminal ratings. Observe all markings on the oscilloscope before connecting. Consult the oscilloscope product manual for further ratings information. Do not remove probe casing. Touching exposed connections may result in electric shock. Use indoors only within operational environment listed.
PP019/PP020 Passive Probe Disposal The probe is marked with this symbol to indicate that it complies with the applicable European Union requirements to Directives 2002/96/EC and 2006/66/EC on Waste Electrical and Electronic Equipment (WEEE) and Batteries. This electronic product is subject to disposal and recycling regulations that vary by country and region. Many countries prohibit the disposal of waste electronic equipment in standard waste receptacles.
Instruction Manual Probe Compensation Proper compensation of the probe is required to assure amplitude accuracy of the waveform being measured by matching the probe to the oscilloscope’s input capacitance. Compensation should be adjusted whenever the probe is connected to or transferred between oscilloscopes. Low Frequency (PP019 and PP020) 1. Apply a 1 kHz square wave to the probe or connect to the oscilloscope’s calibrator output. 2.
PP019/PP020 Passive Probe High Frequency (PP020 only) High frequency should rarely require adjustment. However, if adjustment is required: 1. Apply a 1 MHz square wave to the probe (< 0.7 ns rise-time). 2. Remove the plastic caps from the two, high-frequency trimmers on the PP020 BNC Compensation Box. 3. Adjust HF2 then HF1 until you achieve a flat-topped square wave like the center illustration in the second figure below.
Instruction Manual PP019 Input Impedance Profile PP020 Input Impedance Profile 923445-00 Rev A 9
PP019/PP020 Passive Probe Voltage vs.
Instruction Manual Replaceable Parts Item 1 2 3 4 5 6 7 8 9 Description Sprung Hook, 5 mm, Black GND Lead w/Alligator Clip BNC Adapter, 5 mm Replacement Tip, Black Deluxe Trimmer Tool IC Tip Insulator, 5 mm, Black Tip Insulator, 5 mm, Black Identifier Rings Probe Tip GND Quantity 1 1 1 1 1 1 1 1 1 Standard Accessory Replacement Kit part number is PKIT4-5MM-101.
PP019/PP020 Passive Probe Contact Teledyne LeCroy Teledyne LeCroy Offices World Wide United States and Canada - World Wide Corporate Office Teledyne LeCroy Corporation 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Ph: 800-553-2769 (Sales 1, Applications 2, Service 3) or 845-425-2000 Fax: 845-578-5985 contact.corp@teledynelecroy.com teledynelecroy.
Instruction Manual Teledyne LeCroy Offices World Wide Japan Teledyne LeCroy Japan Hobunsya Funchu Bldg, 3F 3-11-5, Midori-cho, Fuchu-Shi Tokyo, 183-0006 Japan Ph: ++ 81 4 2402 9400 Fax: ++ 81 4 2402 9586 teledynelecroy.com/japan Korea Teledyne LeCroy Korea 10th fl. Ildong Bldg. 968-5 Daechi-dong, Gangnam-gu Seoul 135-280, Korea Ph: ++82 2 3452 0400 Fax: ++82 2 3452 0490 teledynelecroy.com.kr Singapore (Oscilloscopes) Teledyne LeCroy Singapore Pte Ltd.