User Manual
Instruction Manual
Specifications
Model Attn.
Input
Imped.
B/W
Rise
Time
Length
Compensation
Range
R (MΩ)
C (pF)
(MHz)
(ns)
(m)
(pF)
PP019
10x
10
12
250
1.4
1.2
10 - 30
PP020
10x
10
11
500
0.7
1.2
8 - 20
Attenuation Ratio .....................10:1 ± 1%
Max. Input Voltage ...................Measurement Category I: 500 Vrms CAT I
1500 V transient overvoltage
(see voltage derating curve on page 10)
Measurement Category II: 400 Vrms CAT II
Pollution Degree ......................Pollution Degree 2 (as defined in EN 61010-031)
Operating Altitude ....................UP to 3000 meters
Max. Operating Temp ...............0° C to +50° C
Humidity ...................................5% to 95% RH (10° C to 30° C)
5% to 75% RH (30° C to 40° C)
5% to 50% RH (above 40° C)
RH not controlled below 10° C
Definitions:
Measurement Category I (CAT I): measurements performed on circuits not directly connected to mains.
Measurement Category II (CAT II): measurements performed on circuits directly connected to the
low-voltage installation.
Pollution Degree 2: operating environment where normally only dry, non-conductive pollution occurs.
Temporary conductivity caused by condensation must be expected.
923445-00 Rev A 3