Specifications
Model Attn.
Input
Imped.
B/W
Rise
Time
Length
Compensation
Range
R (MΩ)
C (pF)
(MHz)
(ps)
(m)
(pF)
PP025
10x
10
10
500
700
1.3
14 - 25
Attenuation Ratio ..................... 10:1 ± 1%
Max. Input Voltage................... Measurement Category I
1
: 500 Vrms
1500 V transient overvoltage
(see voltage derating curve on p. 3)
Measurement Category II
2
: 400 Vrms
Pollution Degree ...................... Pollution Degree 2
3
Operating Temperature .......... 0° C to +50° C
Humidity .................................... 5% to 95% RH (10° C to 30° C)
5% to 75% RH (30° C to 40° C)
5% to 50% RH (above 40° C)
RH not controlled below 10° C
Operating Altitude .................... 3000 meters maximum
To reduce risk of fire or electric shock, do not exceed the
voltage or category ratings of the probe.
Definitions (per EN 61010-031)
1) Measurement Category I (CAT I) refers to measurements performed on circuits not directly
connected to mains.
2) Measurement Category II (CAT II) refers to measurements performed on circuits directly
connected to the low-voltage installation.
3) Pollution Degree 2 refers to operating environment where normally only dry, non-conductive pollution
occurs. Temporary conductivity caused by condensation must be expected.