Operator’s Manual PP026-1 PP026-2 PP026 Passive Probe
700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com PP026 Passive Probe Instruction Manual © 2015 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales and distribution purposes is strictly prohibited. However, clients are encouraged to duplicate and distribute Teledyne LeCroy documentation for internal educational purposes.
Operator’s Manual Terms and Symbols The following symbols appear on the product or in its documentation: DANGER High Voltage WARNING Refer to Manual. Condition or practice could result in injury or loss of life.
PP026 Passive Probe Specifications Model PP026 Input Imped. Attn. 10x B/W Rise Time Length Compensation Range R (MΩ) C (pF) (MHz) (ps) (m) (pF) 10 10 500 700 1.3 10 - 20 Attenuation Ratio ..................... 10:1 ± 1% 1 Max. Input Voltage................... Measurement Category I : 500 Vrms 1500 V transient overvoltage (see voltage derating curve on p. 3) 2 Measurement Category II : 400 Vrms Pollution Degree ...................... Pollution Degree 2 3 Operating Temperature ......
Operator’s Manual PP026 Input Impedance Profile Voltage vs.
PP026 Passive Probe EC Declaration of Conformity The probe is marked with this symbol to show compliance was demonstrated to the specifications listed in the Official Journal of the European Communities: Low Voltage Directive 2006/95/EC: EN 61010-031/A1:2008 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 031: Safety requirements for handheld probe assemblies for electrical measurement and test.
Operator’s Manual Do not overload. Do not apply any potential to the probe leads that exceeds the maximum rating of the probe. Always comply with the Voltage vs. Frequency Derating Curve. Observe all terminal ratings. Observe all markings on the oscilloscope before connecting. Consult the oscilloscope product manual for further ratings information. Do not remove probe casing. Touching exposed connections may result in electric shock. Use indoors only within operational environment listed.
PP026 Passive Probe Probe Compensation Proper compensation of the probe is required to assure amplitude accuracy of the waveform being measured by matching the probe to the oscilloscope’s input capacitance. Compensation should be adjusted whenever the probe is connected to or transferred between oscilloscopes. Low Frequency 1. Apply a 1 kHz square wave to the probe or connect to the oscilloscope’s calibrator output. 2.
Operator’s Manual Over-compensated, correct, and under-compensated waveforms Service Refer all service to qualified personnel. For the most complete and up-to-date list of sales and service centers by country, visit teledynelecroy.com/support/contact.
PP026 Passive Probe Replaceable Parts Item Part Number 1 2 3 4 5 6 7 8 9 10 11 12 PK6-5MM-101 PK5-PK6-001 PK6-5MM-102 PK5-PK6-002 PK6-5MM-103 PK6-5MM-104 PK6-5MM-105 PK6-5MM-106 PK6-5MM-107 PK6-5MM-108 PK6-5MM-109 PK6-5MM-110 Description Quantity Rigid Probe Tip (1 installed standard) Trimmer Tool Sprung Hook Color Coding Rings (2 ea.
Operator’s Manual Warranty NOTE: THE WARRANTY BELOW REPLACES ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. TELEDYNE LECROY SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT OR OTHERWISE. THE CUSTOMER IS RESPONSIBLE FOR THE TRANSPORTATION AND INSURANCE CHARGES FOR THE RETURN OF PRODUCTS TO THE SERVICE FACILITY.
925647-00 May, 2015