Data Sheet

8
SPECIFICATIONS
Acquire System
Sampling Rate 2GSa/s(single-channel
1)
),1GSa/s(dual-channel)
Memory Depth 140Mpts(single-channel),70Mpts(dual-channel)
Peak Detect 1 ns
Average Averages:4,16,32,64,128,256,512,1024
Eres Enhancebits:0.5,1,1.5,2.,2.5,3selectable
Interpolation Sinx/x, Linear
Input
Channels 2/4 + EXT
Coupling DC, AC, GND
Impedance DC:(1MΩ±2 %)||(22pF±3pF)
50Ω:50Ω±2 %
Max.Input voltage 1MΩ≤400Vpk(DC+PeakAC<=10kHz)
50Ω≤5Vrms
CH to CH Isolation DC~MaxBW>35dB
Probe Attenuation 0.1X,0.2X,0.5X,1X,2X,5X,10X,20X,50X,100X,200X,500X,1000X,2000X,5000X,10000X
Horizontal System
Time Scale 1.0ns/div~50s/div
Channel Skew <100ps
Waveform Capture Rate Upto140,000wfm/s(normalmode),500,000wfm/s(sequencemode)
Intensity grading 256-level
Display Format Y-T,X-Y,Roll
Time base Accuracy ±25ppm
Roll Mode 50ms/div~50s/div(1-2-5Step)
Vertical System
Bandwidth(-3dB) 300 MHz (SDS2304X / SDS2302X)
200 MHz (SDS2204X / SDS2202X)
100 MHz (SDS2104X / SDS2102X)
70 MHz (SDS2074X / SDS2072X)
VerticalResolution 8 bit
VerticalRange 8 divisions
VerticalScale(Probe1X) 1mV/div–10V/div(1-2-5step)
Offset Range (Probe 1X) 1mV/div~100mV/div:±1V
102mV/div~1V/div:±10V
1.02V/div~10V/div:±100V
Bandwidth Limit 20MHz±40 %
Bandwidth Flatness DC~10 %(BW):±1dB
10 %~50 %(BW):±2dB
50 %~100 %(BW):+2dB/–3dB
Low Frequency Response
(AC Coupling –3 dB)
≤10Hz(atinputBNC)
Noise stdev≤0.2div(<2mV/div)
stdev≤0.5div(≥2mV/div)
DC Gain Accuracy 5mV/div~10V/div:≤3.0 %
≤2mV/div:≤4.0 %
Offset Accuracy ≥2mV/div:±(1 %*offset+1.5 %*8*div+2mV)
<2mV/div:±(1 %*offset+1.5 %*8*div+1mV)
Rise Time
1)
(Typ.)<1.2ns(SDS2304X/SDS2302X)
(Typ.)<1.7ns(SDS2204X/SDS2202X)
(Typ.)<3.5ns(SDS2104X/SDS2102X)
(Typ.)<5.0ns(SDS2074X/SDS2072X)
Overshoot(500psRiseEdge) <10 %
All specications are not guaranteed unless the following conditions are met:
The oscilloscope calibration period is valid
Theoscilloscopehasbeenworkingcontinuouslyforatleast30minutesatthespeciedtemperature(18 ~ 28)