CURRENT PROBES
Ordering Information
CA10 Current Sensor Adapter
The CA10 enables a third-party current measurement device to operate like a Teledyne
LeCroy probe. The CA10 is programmable and customizable to work with third-party
current measurement devices that output voltage or current signals proportional to
measured current. (See page 48 for more information and specications).
Specications
CP030
(CP030-3M)
CP030A CP031 CP031A
CP150
(CP150-6M)
CP500
Electrical Characteristics*
Max. Continuous
Input Current
30 A
rms
150 A
rms
500 A
rms
Bandwidth 50 MHz
(10 MHz)
100 MHz
10 MHz
(5 MHz)
2 MHz
Rise Time (typical) ≤ 7 ns
(≤ 35 ns)
3.5 ns
≤ 35 ns
(≤ 70 ns)
175 ns
Max. Peak Current
50 A
peak
(non-continuous)
300 A
peak
(non-continuous);
500 Apeak ≤ 30 μs
700 A
peak
(non-continuous)
Output Voltage 0.1 V/A 0.1 V/A & 1 V/A 0.1 V/A 0.1 V/A & 1 V/A 0.01 V/A
Max Continuous Input Current
at 1 V/A (100mA/div or less)
5 A 5 A
Offset Range at 1V/A
(100mA/div or less)
±5 A ±5 A
Minimum Sensitivity 10 mA/div 1 mA/div 10 mA/div 1 mA/div 100 mA/div
Low-Frequency Accuracy 1%
AC Noise at 20 MHz BWL
2.5 mA 150 µA 2.5 mA 150 µA 6.0 mA 8.0 mA
Coupling AC, DC, GND
General Characteristics
Cable Length 1.5 m
(3 m)
1.5 m
2 m
(6 m)
6 m
Weight 240 g
(290 g)
260 g 240 g 260 g
500 g
(600 g)
630 g
Max. Conductor Size
(Diameter)
5 mm 20 mm
Interface ProBus, 1 M
Ω only
Usage Environment Indoor
Operating Temperature
0
˚ C to 40˚ C
Max. Relative Humidity 80%
Max. Altitude 2000 m
Maximum Insulated
Wire Voltage
300 V CAT I 600 V CAT II, 300 V CAT III
* Electrical Characteristics Guaranteed at 23 °C ±3 °C
CP03x, CP150, and CP500 probes (and long cable versions of these) are compatible with any Teledyne LeCroy oscilloscope with a ProBus interface running
rmware version 4.3.1.1 or greater.
CP03xA probes are compatible with most Teledyne LeCroy oscilloscopes with a ProBus interface running X-Stream™ rmware version 7.8.x.x or later.
Product Description Product Code
ProBus Current Sensor Adapter CA10
Set of 4 CA10, ProBus Current Sensor Adapters CA10-QUADPAK
30 A; 50 MHz Current Probe – AC/DC; 30 A
rms
; 50 A Peak Pulse, 1.5 meter cable CP030
30A; 10 MHz Current Probe - AC/DC, 30 Arms; 50 A Peak Pulse, 3 meter cable (not EMC compliant) CP030-3M
30 A; 50 MHz High Sensitivity Current Probe – AC/DC; 30 A
rms
; 50 A Peak Pulse, 1.5 meter cable CP030A
30 A; 100 MHz Current Probe – AC/DC; 30 A
rms
; 50 A Peak Pulse, 1.5 meter cable CP031
30 A; 100 MHz High Sensitivity Current Probe – AC/DC; 30 A
rms
; 50 A Peak Pulse, 1.5 meter cable CP031A
150 A; 10 MHz Current Probe – AC/DC; 150 A
rms
; 500 A Peak Pulse, 2 meter cable CP150
150 A; 5 MHz Current Probe – AC/DC; 150 Arms; 500 A Peak Pulse, 6 meter cable (not EMC compliant) CP150-6M
500 A; 2 MHz Current Probe – AC/DC; 500 A
rms
; 700 A Peak Pulse, 6 meter cable CP500
Deskew Calibration Source for CP030, CP030A, CP031, CP031A, AP015, CP150, CP500 DCS025