Data Sheet

39
HIGH VOLTAGE FIBER OPTICALLY-ISOLATED PROBES
Product Description Product Code
High Voltage Fiber Optically-isolated Probe Models and Accessories
High Voltage Fiber Optic Probe, 60 MHz Bandwidth. Includes soft-carrying case, Qty. 1 Amplier/Modulating Transmitter, Qty. 1
Demodulating Receiver, Qty. 1 1m Fiber Optic Cable, Qty. 1 USB Charging Cable, Qty. 1 Micro-gripper set.
Attenuating Tips must be ordered separately.
HVFO103
HVFO10X +/-1V (1x Attenuation) Universal Tip Accessory HVFO100-1X-TIP-U
HVFO10X +/-5V (5x Attenuation) Universal Tip Accessory HVFO100-5X-TIP-U
HVFO10X +/-10V (10x Attenuation) Universal Tip Accessory HVFO100-10X-TIP-U
HVFO10X +/-20V (20x Attenuation) Universal Tip Accessory HVFO100-20X-TIP-U
HVFO10X +/-40V (40x Attenuation) Universal Tip Accessory HVFO100-40X-TIP-U
Spare Amplier/Modulating Transmitter (permits constant operation - charge one while another is being used) HVFO103-XMITTER
1m Spare Fiber Optic Cable Accessory HVFO-1M-FIBER
2m Fiber Optic Cable Accessory HVFO-2M-FIBER
6m Fiber Optic Cable Accessory HVFO-6M-FIBER
NIST Traceable Calibration Certicate HVFO103-CCNIST
Electrical
Bandwidth 60 MHz (typical, with tip attached)
Rise Time (10-90%) 7.5 ns (typical)
Input Dynamic Range ±1V, ±5V, ±10V, ±20V, ±40V (DC+peak AC) respectively with 1X, 5X, 10X, 20X or 40X attenuating tips.
All tips are purchased as accessories (none are included with HVFO103 probe).
Maximum Non-destruct Voltage 5 times the operating voltage rating (tip dependent)
Common Mode Voltage Range ±35 kV (DC+Peak AC) (not for hand-held use, with adequate spacing between probe components and earth ground)
Maximum Input Voltage to Earth ±35 kV (DC+Peak AC) (not for hand-held use, with adequate spacing between probe components and earth ground)
Maximum Safe Input Voltage For hand-held use, 30 Vrms / 60 Vdc per IEC/EN 61010-031:2015
Offset Offset capability determined by oscilloscope offset available in a given gain (V/div) setting after accounting for total
probe attenuation (total probe attenuation is twice the tip attenuation).
Sensitivity 50 mV/div to 1 V/div (1X tip), 250 mV/div to 5 V/div (5X tip), 500 mV/div to 10 V/div (10X tip), 1 V/div to 20 V/div (20X tip),
2 V/div to 40 V/div (40X tip)
Gain Accuracy 2% (LF, guaranteed)
Input Impedance 1 MΩ || 34 pF (1X tip); 5 MΩ || 26 pF (5X tip); 8 MΩ || 23 pF (10X tip); 10 MΩ || 22 pF (20X tip); 10 MΩ || 22 pF (40X tip)
Input/Output Coupling DC only
Interface ProBus
Cable Length 1.25 m (4.1 feet) from input lead to oscilloscope connection (using included 1 meter ber optic cable)
Battery 6 hour battery life (typical). 1.5 hour re-charge time (typical, with user-supplied dedicated USB charger).
3 hour re-charge time (typical) using supplied USB charging cable connected to oscilloscope USB port
Noise, Rejection, and Electromagnetic Compatibility (EMC)
CMRR (typical) 140 dB (100 Hz), 120 dB (to 1 MHz), 85 dB (to 10 MHz),
60 dB (to 60 MHz)
Noise (Probe only) 7 mVrms (1X tip), 35 mVrms (5X tip), 70 mVrms (10X tip), 140 mVrms (20X tip), 280 mVrms (40X tip)
Electrostatic Discharge (ESD)
Immunity
8 kV contact discharge and 10 kV air discharge per IEC61000-4-2, criteria A
Radiated RF Electromagnetic
Field Immunity
25 V/m (80 MHz to 2.7 GHz) per IEC61000-4-3, criteria A
Immunity to Conducted
Disturbance Induced by
RF Fields
10 V/m (150 kHz to 80 MHz) per IEC61000-4-6, criteria A
Environmental
Temperature 0°C to 50°C (operating), -20°C to 70°C (non-operating)
Humidity 5% to 80% RH (non-condensing) up to 30°C, decreasing linearly to 45% RH at 50°C (operating)
5% to 95% RH (non-condensing), 80% RH above 30°C, 45% RH above 50°C (non-operating)
Altitude Up to 3000 m (operating), 10,000 m (non-operating)
Pollution Degree 2, Indoor Use Only
Certications
CE Declaration of Conformity Low Voltage Directive 2014/35/EU (IEC/EN 61010-031:2015
EMC Directive 2014/30/EU (IEC/EN 61326-1:2013)
RoHS2 Directive 2011/65/EU
Ordering Information