Instructions

T3RC3000-LF AND T3RC6000-LF HANDLING INSTRUCTIONS
The coil free end is fully inserted in the ferrule when the user feels the free end of the coil engage
with the internal click-in mechanism inside the ferrule.
Visual indication that the coil mechanism is fully inserted is provided by the black coil marker
situated near the end of the coil, as shown below:
Once the free end of the coil is fully engaged in the ferrule, turn the ferrule lock nut to lock the coil
end in position.
To remove the probe from the device under test rst loosen the ferrule lock nut then remove the
coil end from the ferrule.
The locking mechanism must not be actuated at temperatures below 0 C.
Care must be taken not to over-tighten the locking nut.
When not in use return the Rogowski probe to its protective case.
T3RC3000-LF and T3RC6000-LF SAFETY AND PRE-USE
CHECKS
The integrity of the insulation around the Rogowski coil itself must be VISUALLY
INSPECTED before use, and the T3RC3000-LF and T36000-LF must NOT BE
USED if there are signs of damage.
When bending the exible coil around a conductor, avoid tight bends and sharp edges
that could damage the coil.
11