Instructions
3. Insert the ‘free end’ of the coil fully inside the ferrule.
4. Re-energise the conductor.
5. Use the push button to switch ON by pressing and releasing the button, and to turn OFF
by depressing the button fully; the LED indicates that the transducer is ON when the LED
is GREEN.
6. After switch-on the Teledyne Test Tools Rogowski probes, they may require a settling period
of up to 2 minutes before providing a correct current measurement. The settling period is
dependent on warm-up time and the low frequency bandwidth.
OBTAINING THE BEST MEASUREMENT
The Rogowski coil should be positioned so that the conductor under test is encircled by the coil
but is not adjacent to the cable attachment (see picture below). The diagram shows the direction
a positive current should pass through the coil loop in order to obtain a positive output voltage.
The Teledyne Test Tools Rogowski probe has been calibrated with the conductor near the
centre (position A), and this is the ideal position for accuracy.
For the best high frequency performance, the centre of the current should lie on the line shown
A to B, where B is half way around the circumference of the coil.
SENSITIVITY TO EXTERNAL CURRENTS AND VOLTAGES
The sensitivity of the Teledyne Test Tools Rogowski probes to currents that are outside the closed
loop of the Rogowski coil is very small, provided that the external currents are less than the
current rating of the Teledyne Test Tools Rogowski probes or that such currents are relatively
distant from the coil. In the vicinity of a multi-turn conductor the eects are far stronger than from
a single conductor carrying the same current, and such positions should be avoided.
I current
into page
KEEP CONDUCTORS
AWAY FROM
SHADED AREA