Owner manual

TA660 User's Manual Catalyst Enterprises, Inc
24
Data Capture Options
Memory Display You may limit the captured data display to a specific
number of samples by checking Partial Memory and
entering the number of Samples to be captured or, you may
check entire memory to allow the capture for the entire
memory.
Pre-Trigger Pre-Trigger is set by default at 50% which defines the
percentage of data to be captured before and after the
triggering event. You may change this percentage by
dragging the slider to the desired value.
Pre-Trigger Data: The capture of the specified percentage of the data prior to
the triggering event cannot be guaranteed and may in some
cases be 0. This can occur in cases where the triggering
event occurs before the required number pre-trigger event
data can be stored. In these cases the data display will show
fewer than the specified data points prior to the triggering
event. For more detail see Set Trigger on page 69.
Manual Trigger Select this option when you wish to manually interrupt the
data capture based on some external event. With this option
selected, data will be continually captured to memory and
overwritten as required until you stop the data capture.
Continuous Mode Continually captures data for the number of times specified.
A separate data file is generated each time that a capture is
performed.
Exclude Cycles To simplify the captured data display you may check the
options to exclude, Idles, Wait States and Retries.