Data Sheet

Table Of Contents
12
1 INTRODUCTION
Integrated in Teledyne LeCroy’s Voyager M310C test platform, the Power Delivery
exerciser supports traffic generation, including both provider and consumer device
emulation. The Power Delivery exerciser continues to evolve with each software release.
Be sure to check for updated software and firmware before getting started with the
Exerciser.
Important Licensing Note:
Operating the PD Exerciser beta requires that the USB Power Delivery
Exerciser option is enabled on the M310C base unit:
Getting Started:
The “left” port of the Voyager should be used to
connect DUT to the PD Exerciser. The PD exerciser
also requires specific cable orientation (Red LED when
connected wrong side-up).
To enable the PD Trainer/Exerciser, use the PD Tab under “Recording Options”
to select the Exerciser mode.
Note Allow VBUS > 5v is a safety feature which prevents sourcing above 5V. When
enabled, this mode will allow Voltage levels to be delivered to the DUT which may exceed
their current carrying capabilities. While the M310C system is designed to tolerate higher
current, these higher voltages may inadvertently cause damage to devices/cables under
test.
To set devices port name, use the General Tab under “Recording Options” to add
“alias labels” for your DUTs.
These labels will appear in the trace capture.
The Alias name is primarily for use in analyzer mode and requires that device names are
added before recording traffic. The device naming can also be used in Exerciser mode;
however message frames from the Voyager M310C will be always be labeled “M310C”.