User manual

WaveStation SCPI Command Reference Manual
920595 RevA
20
Tables specifically covering command arguments are listed in the order in which they are functionally
processed for convenience.
Where:
<state>:= {ON,OFF}
<src>:= {INT,EXT}
<mod wave shape>:= {SINE,SQUARE,TRIANGLE,UPRAMP,DNRAMP,NOISE,ARB}
<am frequency>:= {0.002Hz to 20000Hz}
<deep>:= {0 to 120}
<fm frequency>:= {0.002Hz to 20000Hz}
<fm frequency offset>:= {0 to basic waveform frequency / 2}
<pm frequency>:= {0.002Hz to 20000Hz}
<pm phase offset>:= {0 to 360}
<pwm frequency>:= {0Hz to 4kHz}
<pwm devi>:= {Deviation is dependent on the carrier waveform duty cycle}
<ask key frequency>:= {0.002Hz to 20000Hz}
<fsk frequency>:= {0.002Hz to 50000Hz}
<fsk hop frequency>:= {The hop frequency range corresponds to the basic
waveform frequency range}
<wave type>:= {SINE ,SQUARE, RAMP, ARB, PULSE}
<frequency>:= {The default unit is in HZ. The minimum value is 1xe-6 HZ,
maximum value depends on your WaveStation model.}
<amplifier>:= {The default unit is in V. The C1 minimum value is 0.004V,
Maximum is 6V. The C2 minimum value is 0.004V, and its maximum value is
20V.}
<offset>:= {The default unit is in V.}
<duty>:= {If waveform type is set to Square, the range is from 20% to 80%. If the
waveform type is set to Pulse, the range is from 0.1% to 99.9%.}
<symmetry>:= {0% to 100%}
<delay>:= {The maximum value is 2 Ks}
QUERY SYNTAX
<channel>: MoDulateWaVe?
<channel>:= {C1,C2}
RESPONSE FORMAT
<channel>: MoDulateWaVe <parameter>
<parameter>:= {Return all parameters for the current modulated waveform.}
EXAMPLE 1
Set the C1 modulation type to AM.
C1: MDWV AM