Manual

Remote Control Commands and Queries
WM-RCM-E Rev D ISSUED: February 2005
115
N
N
O
O
T
T
E
E
:
: The keyword STAT is optional with modes CUST, HPAR, and VPAR. If present, STAT turns
parameter statistics on. Absence of STAT turns parameter statistics off.
The submodes ABS, DELTA, SLOPE are optional with mode HREL. If it is present, ABS
chooses absolute amplitude reading of relative cursors. Absence of keyword selects relative,
DELTA, amplitude reading of relative cursors. Submodes ABS and DELTA are used with
VREL; presence of ABS selects absolute amplitude, absence selects DELTA.
QUERY SYNTAX
CuRsor_MeaSure?
RESPONSE FORMAT
CuRsor_MeaSure <mode>
EXAMPLE (GPIB) The following instruction switches on the vertical relative cursors:
CMD$=“CRMS VREL”: CALL IBWRT(SCOPE%,CMD$)
The following instruction determines which cursor is currently turned
on:
CMDS$=“CRMS?”: CALL IBWRT(SCOPE%,CMD$):
CALL IBRD(SCOPE%,RD$): PRINT RD$
Example of response message:
CRMS OFF
RELATED COMMANDS
CURSOR_SET, PARAMETER_STATISTICS,
PARAMETER_VALUE, CURSORS, PARAMETER,
PASS_FAIL
ADDITIONAL INFORMATION To turn off the cursors, parameter measurements or Pass/Fail tests,
use:
CURSOR_MEASURE OFF