CURSOR
CURSOR_SET, CRST
Command/Query
DESCRIPTION
The CURSOR_SET command allows you to position any one of
the independent cursors at a given grid location.
When you are setting a cursor position, you must specify a trace,
relative to which the cursor will be positioned. This means that
the trace must be turned on, a requirement that does not apply to
all earlier LeCroy instruments. Also new to X-Stream
instruments is that PREF and PDIF are not supported, because
each parameter has its own independent gate.
N
N
O
O
T
T
E
E
:
: To change only the trace without repositioning the
cursors, the CURSOR_SET command can be given with no
argument (for example, F2:CRST).
The CURSOR_SET? query indicates the current position of
the cursor(s). The values returned depend on the grid type
selected.
N
OTATION
HABS
horizontal absolute
VABS
vertical absolute
HDIF
horizontal difference
VDIF
vertical difference
HREF
horizontal reference
VREF
vertical reference
COMMAND SYNTAX
<trace> : CuRsor_SeT <cursor>,<position>[,<cursor>,<position>,
<cursor> ,<position>]
<trace> : = {F1, F2, F3, F4, F5, F6, F7, F8, TA, TB, TC, TD, C1,
C2, C3, C4,M1,M2,M3,M4} TA through TD are for compatibility
with existing software with earlier instruments. These four
mnemonics are not returned by queries.
<cursor> : = {HABS, VABS, HREF, HDIF, VREF, VDIF}
<position> : = 0 to 10 DIV (horizontal)
<position> : = -3.99 to 3.99 DIV (vertical)