Manual

Remote Control Commands and Queries
WM-RCM-E Rev D ISSUED: February 2005
155
N
N
O
O
T
T
E
E
:
: The segment number is only relevant for waveforms
acquired in sequence mode; it is ignored in single waveform
acquisitions. When the segment number is set to 0, all segments
will be shown.
The unit DIV is optional.
QUERY SYNTAX
<exp_trace> : Hor_POSition?
RESPONSE FORMAT
<exp_trace> : Hor_POSition <hor_position>[,<segment>]
N
N
O
O
T
T
E
E
:
: The segment number is only given for sequence
waveforms.
EXAMPLE (GPIB) The following instruction positions the center of the intensified zone
on the trace currently viewed by Trace A (TA) at division 3:
CMD$=“F1:HPOS 3”: CALL IBWRT(SCOPE%,CMD$)