Remote Control Commands and Queries
WM-RCM-E Rev D ISSUED: February 2005
219
SAVE/RECALL
SETUP
*SAV
Command
DESCRIPTION
The *SAV command stores the current state of your instrument in
non-volatile internal memory. The *SAV command stores the
complete front panel setup of the oscilloscope at the time the
command is issued.
N
N
O
O
T
T
E
E
:
: Neither communication parameters (those modified
by the commands
COMM_FORMAT
,
COMM_HEADER
,
COMM_HELP
,
COMM_ORDER,
and
WAVEFORM_SETUP
) nor
enable registers of the status reporting system (
*SRE
,
*PRE
,
*ESE
,
INE
) are saved when
*SAV
is used.
COMMAND SYNTAX
*SAV <panel_setup>
<panel_setup> : = 1 to 6
EXAMPLE (GPIB) The following instruction saves the current instrument setup in
panel setup 3:
CMD$=“*SAV 3”: CALL IBWRT(SCOPE%,CMD$)
RELATED COMMANDS
PANEL_SETUP, *RCL