Manual

P
ART
T
WO
:
COMMANDS
226
ISSUED: February 2005
WM-RCM-E Rev D
ACQUISITION
STOP
Command
DESCRIPTION
The STOP command immediately stops the acquisition of a signal.
If the trigger mode is AUTO or NORM, STOP will place the
oscilloscope in STOPPED trigger mode to prevent further
acquisition.
COMMAND SYNTAX
STOP
EXAMPLE The following instruction stops the acquisition process:
CMD$ =“STOP”: CALL IBWRT(SCOPE%,CMD$)
RELATED COMMANDS
ARM_ACQUISITION, TRIG_MODE, WAIT,
FORCE_TRIGGER