Manual

Remote Control Commands and Queries
WM-RCM-E Rev D ISSUED: February 2005
245
INTV
Interval
TEQ
Edge Qualified
IS
Interval smaller than
TI
Time greater than
I2
Interval width
TL
Time within
COMMAND SYNTAX
TRig_SElect <trig_type>,SR,<source>,QL,<source>,
HT,<hold_type>,HV,<hold_value>,HV2,<hold value>
<trig_type> : = {DROP, EDGE, GLIT, INTV, STD, SNG,
SQ, TEQ}
<source> : =
{
C1, C2, C3, C4, LINE, EX, EX10, PA,
ETM10}
<hold_type> : = {TI, TL, EV, PS, PL, IS, IL, P2,
I2, OFF}
<hold_value> : = See specifications in the Operator’s Manual for valid
values
N
N
O
O
T
T
E
E
:
: The unit S (seconds) is optional.
QUERY SYNTAX
TRig_SElect?
RESPONSE FORMAT
TRig_SElect
<trig_type>,SR,<source>,HT,<hold_type>,HV,
<hold_value>,<hold_value>
¾ HV2 only returned if <hold_type> is P2 or I2