Manual

P
ART
T
WO
:
COMMANDS
258
ISSUED: February 2005
WM-RCM-E Rev D
Segment number (SN): The segment number parameter indicates which segment should be
sent if the waveform was acquired in sequence mode. This parameter
is ignored for non-segmented waveforms. For example:
SN = 0
all segments
SN = 1
first segment
SN = 23
segment 23
The WAVEFORM_SETUP? query returns the transfer parameters
currently in use.
COMMAND SYNTAX
WaveForm_SetUp
SP,<sparsing>,NP,<number>,FP,<point>,SN,<segment>
N
N
O
O
T
T
E
E
:
: After power-on, all values are set to 0 (i.e., entire waveforms will be transmitted without
sparsing).
Parameters are grouped in pairs. The first of the pair names the variable to be modified, while the
second gives the new value to be assigned. Pairs can be given in any order and restricted to those
variables to be changed.
QUERY SYNTAX
WaveForm_SetUp?
RESPONSE FORMAT
WaveForm_SetUp
SP,<sparsing>,NP,<number>,FP,<point>,SN, <segment>
EXAMPLE (GPIB) The following instructs every 3rd data point (SP=3) starting at
address 200 to be transferred:
CMD$=“WFSU SP,3,FP,200”: CALL
IBWRT(SCOPE%,CMD$)
RELATED COMMANDS
INSPECT, WAVEFORM, TEMPLATE
§ § §