Returns: True on success, False on failure.
Remarks: This method sends a string command to the instrument.
If EOI is set to TRUE, the device will start to interpret the command immediately. This is normally the
desired behavior.
If EOI is set to FALSE, a command may be sent in several parts with the device starting to interpret the
command only when it receives the final part, which should have EOI set to TRUE.
§ § §