Datasheet

14
1.5 GHz DIFFERENTIAL PROBES
The ZD Series probes provide wide dynamic range, excellent noise and loading
performance and an extensive set of probe tips, leads, and ground accessories
to handle a wide range of probing scenarios. The low 1 pF capacitance means
this probe is ideal for all frequencies. The ZD Series differential probes provide full
system bandwidth for all Teledyne LeCroy Oscilloscopes 1.5 GHz and lower.
Fully Integrated
With the ProBus interface, the ZD500, 1000, and 1500 become an integral part
of the oscilloscope. All probe gain and offset controls are transparent to the user,
making it easier to probe the circuit without concern for which gain setting to
choose. When used with a Teledyne LeCroy digital oscilloscope, no external power
supply is required.
Wide Dynamic Range
The ZD500, 1000, 1500 probes provide transparent probe attenuation so signals are
always optimized for the display. The differential range is 18 V
p-p
with a differential
offset of ±8V and common mode range of ±10 V, making these probes versatile for
every probing application.
Wide Applications
The wide dynamic range of 16 V
p-p
and offset range of ±8V suit this probe to a wide
range of applications and signal types. The ZD differential probes are ideally suited
for Automotive, Serial Data, power, and general purpose use.
Specications
ZD200 ZD500 ZD1000 ZD1500
Electrical Characteristics
Bandwidth (Warranted) 200 MHz 500 MHz 1000 MHz 1500 MHz
Bandwidth (Typical)
-
650 MHz 1200 MHz 1700 MHz
Risetime 10–90% (Typical) 1.75 ns 650 ps 375 ps 270 ps
Risetime 20–80% (Typical)
-
500 ps 280 ps 200 ps
LF Attenuation Accuracy (Warranted) 1% 2%
Zero Offset (Typical)
(within 15 minutes after autozero)
-
5 mV
System Noise (Typical)
-
1.3 mVrms 1.75 mVrms
Probe Noise Density (Typical) 3 mV
rms
38 nV/rt (Hz)
Input Differential Range (Nominal) ± 20 V ±8 V (16 V
p-p)
Differential Offset Range (Nominal)
-
±18 V
Offset Gain Accuracy (Typical)
-
2%
Common Mode Range (Nominal) ± 60 V ±10 V
Maximum Non-destruct Voltage
(Nominal)
-
30 V
CMRR (Typical) 80 dB @ 60 Hz
50 dB@10 MHz
60 dB 50/60 Hz
30 dB 20 MHz
25 dB 500 MHz
60 dB 50/60 Hz
30 dB 20 MHz
25 dB @ 1000 MHz
60 dB 50/60 Hz
30 dB 20 MHz
25 dB @ 1500 MHz
DC Input Resistance (Nominal) 250 kΩ (Common Mode)
1 MΩ (Differential Mode)
50 kΩ (Common Mode)
120 kΩ (Differential Mode)
Differential Input Capacitance
(Typical)
3.5 pF < 1.0 pF
Teledyne LeCroy
1.5 GHz
Differential Probe
Model Numbers:
ZD200
ZD500
ZD1000
ZD1500
AP033