Datasheet
34
HIGH VOLTAGE PROBES
Teledyne LeCroy
High Voltage Probe
Model Number:
HVP120
Electrical Characteristics
Bandwidth 400 MHz
Risetime (10% - 90%) 900 ps (typical)
Maximum Input Voltage*
Measurement Category II 1000 Vrms
Measurement Category I 4000V Transient Overvoltage at 1000 Vrms
6000V Transient Overvoltage at 0 Vrms
Pollution Degree* 2
Input Capacitance 7.5 pF (typical)
Compensation Range 10 pF - 50 pF (typical)
Attenuation Ratio 100:1 ± 2%
Environmental
Temperature (Operating) 0°C to 50°C
Temperature (Non-Operating) -40°C to 71°C
Humidity (Operating) 80% RH (Non-Condensing) up to 31°C,
decreasing linearly to 40% RH at 50°C
Altitude (Operating) up to 2,000 m
Altitude (Non-Operating) up to 15,000 m
General Characteristics
Weight (probe) 67 g (0.15 lbs)
Cable Length 2 m (6.56 ft)
Probe Tip Diameter 5 mm (0.20 inches)
* As dened in IEC 61010-031
The HVP120 is a high voltage passive probe designed for probing up to
1,000 Vrms and capable of handling up to 6,000 V peak transients. Its
fast rise time and excellent frequency response make it suitable for a wide
variety of high voltage measurement applications. The HVP120 features
a spring loaded probe tip and a variety of standard accessories to make
probing high voltages safe and easy.
Features
•
400 MHz probe bandwidth
•
900 ps rise time
•
1000 Vrms maximum input
•
Up to 6 kV transient overvoltage