Compliance Information
This LeCroy accessory is CE Compliant and bears the CE mark.
CE Declaration of Conformity
This ZD Series Probes meet intent of the European Council Directive 2006/95/EEC for Product
Safety and 2004/108/EEC for Electromagnetic Compatibility. Compliance was demonstrated to
the following specifications as listed in the Official Journal of the European Communities:
EN 61010-031:2002 Safety requirements for electrical equipment for measurement,
control, and laboratory use.
Part 031: Safety requirements for the hand-held probe assemblies for electrical
measurement and test.
The design of the differential probe has been verified to conform to EN 61010-031 safety
standard per the following limits:
o Installation (Overvoltage) Category I: Refers to signal level which is applicable for
equipment measuring terminals that are connected to source circuits in which
measures are taken to limit transient voltages to an appropriate low level.
o Pollution Degree 2: Refers to an operating environment where normally only dry
non-conductive pollution occurs. Occasionally a temporary conductivity caused by
condensation must be expected.
EN 61326-1:2006, EN 61326-2-1:2006 EMC requirements for electrical equipment for
measurement, control and laboratory use.
o EN 55011/A2:2002 Radiated and Conducted Emissions, Group 1, Class A *
o EN 61000-4-2:2001 Electrostatic Discharge Immunity (±4 kV contact discharge, ±4
kV air discharge, ±4 kV vertical/horizontal coupling planes )
o EN 61000-4-3:2006 RF Radiated Electromagnetic Field Immunity (3 V/m, 80 MHz to
1 GHz; 3 V/m, 1.4 GHz to 2 GHz; 1 V/m, 2 GHz to 2.7 GHz , 80% amplitude
modulated with 1 kHz sinewave)
* This product is intended for use in nonresidential areas only. Use in residential areas may
cause electromagnetic interference.