User manual
ZD Series Differential Probes
ZDSERIES-GSM-REVA
34
Equipment Used
Instrument
Model
Serial Number
Calibration Due Date
Oscilloscope
Digital Multimeter
Function Generator
Note: The function generator used in this Performance Verification Procedure is used for
making relative measurements. The output of the generator is measured with a DMM or
oscilloscope in this procedure. Thus, the generator is not required to be calibrated.
Test Record
Output Zero Voltage
Step
Description
Intermediate Data
A-5.
Output Zero (Test limit ≤ ±800 μV)
LF Attenuation Accuracy
Step
Description
Intermediate Data
B-7
Generator Output Voltage
V
B-8
Expected Output Voltage, top range
V
B-10
Measured Output Voltage, top range
V
B-12
Gain Error, top range (Test Limit ≤ ± 1.0%)
%
B-17
Generator Output Voltage
V
B-19
Expected Output Voltage, top range
V
B-21
Measured Output Voltage, top range
V
B-23
Gain Error, top range (Test Limit ≤ ± 1.0%)
%