Instruction M anual ZD Series Differential Probes (ZD500, ZD1000, ZD1500)
ZD Series Differential Probes Instruction Manual March, 2011
LeCroy Corporation 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 578-6020, Fax: (845) 578 5985 Warranty LeCroy warrants this oscilloscope accessory for normal use and operation within specification for a period of one year from the date of shipment. Spare parts, replacement parts and repairs are warranted for 90 days.
ZD Series Differential Probes TABLE OF CONTENTS Introduction ..........................................................................................................................7 Key Benefits .......................................................................................................................7 Standard Accessories..........................................................................................................7 Features and Accessories ........................................
ZD Series Differential Probes Cleaning ........................................................................................................................... 17 Calibration Interval .......................................................................................................... 17 Service Strategy ............................................................................................................... 17 Returning a Probe for Calibration or Service ......................................
ZD Series Differential Probes Test Record ...................................................................................................................... 34 Output Zero Voltage .............................................................................................................. 34 LF Attenuation Accuracy ........................................................................................................
Instruction Manual Introduction The ZD Series of Differential Probes (ZD500, ZD1000 and ZD1500) are high bandwidth active differential probes. The probes feature low noise, very high input impedance and high common mode rejection, and are ideally suited for signal integrity measurements in high-speed digital systems. With low input capacitance and high input resistance, circuit loading is minimized.
ZD Series Differential Probes Standard Accessory Quantity Part Number Mini-Grabber 2 PACC-CL001 Spring-loaded Ground (Long) 2 PACC-ZD003 Spring-loaded Ground (Short) 2 PACC-CD008 Small IC Adapter 2 PACC-ZD006 Tip Saver 2 PACC-ZD004 Y-lead Adapter 1 PACC-ZD001 Swivel Tip Adapter 1 PACC-ZD005 Probe Calibration Fixture 1 PCF200 Instruction Manual 1 Certificate of Calibration 1 ZDSERIES-GSM-REVA 8
Instruction Manual Features and Accessories The ZD Series probes are provided with numerous features and accessories to make probing and connecting to different test points easier than ever. The small, low mass probe head is designed for ease of use and high performance. The probe tip socket fits easily onto 0.025 inch square pins for direct access to test points. Several different adapters are available which connect directly in the probe socket.
ZD Series Differential Probes Tip Saver To prevent excessive wear on the probe input leads, it is recommended to use tip saver in most probing scenarios. The tip saver offers full system bandwidth and will not degrade signal under test. Grabbers Micro- and Mini-Grabbers The micro- and mini-grabbers are ideal for connecting to small IC legs or pins very tightly spaced.
Instruction Manual Grounds Long and Short Spring Loaded Bendable Ground The bendable ground leads are designed to be attached to the offset ground socket or be attached to either socket of the probe head. The following image is an example using the short spring-loaded, bendable ground.
ZD Series Differential Probes Leads While longer leads provide greater flexibility when connecting the probe to a circuit, the added inductance may degrade the fidelity of high frequency signals. Short and Long Right Angle Lead This lead has a socket on one end and a square pin on the other to connect to the input or ground socket of the probe body, and may be used for general purpose probing or can be connected to the Mini-Grabber or Micro-Grabber accessories.
Instruction Manual Y Lead Adapter This lead is used for both ground and input lead simultaneously. It has two sockets on one end and two square pins on the other and may be used for general purpose probing. Resistive compensation to reduce inductive peaking is included. Solder-In Lead This lead can be soldered directly to the test points for a secure probe connection. Small IC Adapter The Small IC adapters are designed for probing the leads of an IC.
ZD Series Differential Probes Probe Calibration Fixture The probe deskew fixture (PCF-200) is provided as a standard accessory with ZD500, ZD1000, and ZD1500 probes. The fixture can be used one of two ways: 1. The fixture may be used to determine the effect of probe input loading on the circuit under test and verification of the probe response to the signal being measured. 2. The fixture may be used as a convenient way to deskew several probes/oscilloscope channels.
Instruction Manual Probe Operation Handling the Probe The ZD Series probes are precision test instruments. Exercise care when handling and storing the probe. Always handle the probe by the probe body or compensation box. Avoid putting excessive strain or exposing the probe cable to sharp bends. ESD Sensitive: The tips of the ZD Series probes are sensitive to Electrostatic Discharge (ESD). Avoid causing damage to the probe by always following anti-static procedures (wear wrist strap, etc.
ZD Series Differential Probes Connecting the Probe to the Test Circuit To maintain the high performance capability of the probe in measurement applications, care must be exercised in connecting the probe to the test circuit. Increasing the parasitic capacitance or inductance in the input paths may introduce a “ring” or slow the rise time of fast signals.
Instruction Manual Care and Maintenance Cleaning The exterior of the probe and cable should be cleaned, using a soft cloth moistened with water. The use of abrasive agents, strong detergents, or other solvents may damage the probe. Always ensure that the input leads are free of debris. The probe case is not sealed and should never be immersed in any fluid. Calibration Interval The recommended calibration interval is one year.
ZD Series Differential Probes Products returned to the factory require a Return Material Authorization (RMA) acquired by contacting your nearest LeCroy sales office, representative or the North America Customer Care Center. Return shipment should be prepaid. LeCroy cannot accept COD or Collect Return shipments. We recommend air-freighting. Note: It is important that the RMA be clearly shown on the outside of the shipping package for prompt redirection to the appropriate department.
Instruction Manual Returning a Probe to a Different Country Note: Be sure to properly mark shipments returned for service from a different country to avoid customs duty for a full purchase price of a new probe or accessory. In addition to the items mentioned in the previous topic, mark shipments returned for service as a Return of US manufactured goods for warranty repair/recalibration.
ZD Series Differential Probes Replacement Parts The probe connection accessories and other common parts can be ordered through the North America Customer Care Centers.
Instruction Manual Performance Verification Performance Verification Overview This procedure can be used to verify the warranted characteristics of the ZD Series High Impedance Active Probe. The recommended calibration interval for the models ZD Series is one year. The complete performance verification procedure should be performed as the first step of annual calibration. Test results can be recorded on a photocopy of the Test Record provided in Appendix A at the end of the manual.
ZD Series Differential Probes Description Digital Oscilloscope Digital Multimeter (DMM) with test probe leads Minimum Requirement ProBus Interface; Windowsbased 4.5 digit DC: 0.1% Accuracy AC: 0.1% Accuracy Test Equipment Examples LeCroy WaveRunner Zi or WaveSurfer Xs Agilent Technologies 34401A or Fluke 8842A-09 Function Generator Sine Wave output amplitude adjustable to 14.
Instruction Manual Preliminary Procedure 1. Connect the ZD Series probe to the female end of the ProBus Extension Cable. Connect the male end of the ProBus Extension Cable to channel 1 of the oscilloscope. 2. Turn the oscilloscope on and allow at least 30 minutes warm-up time for the ZD Series and test equipment before performing the Verification Procedure. 3. Turn on the other test equipment and allow them to warm up for the manufacturer’s recommended timeframe. 4.
ZD Series Differential Probes Functional Check The functional check will verify the basic operation of the probe functions. It is recommended that the Functional Check be performed prior to the Performance Verification Procedure. 1. Return to the factory default settings by: a. Selecting File, Recall Setup... from the menu bar. b. Then touching the Recall Default button. 2. Touch the C1 trace label to open the C1 Vertical Adjust dialog. 3.
Instruction Manual B. LF Attenuation Accuracy 1. Disconnect the BNC tee at the power supply from the dual banana plug adapter. Connect the BNC tee to the output of the function generator. (Use a 50 Ω termination if the function generator requires such a load.) 2. Carefully insert the Straight Tips (supplied in accessory kit) into the sockets of the probe head. Attach the red lead of the mini-grabber to the positive (+) signal input and the black lead to the negative (-) input of the probe head. 3.
ZD Series Differential Probes 16. Adjust the sine wave generator output amplitude to approximately 1 Vrms as measured on the DMM. 17. Record the reading to 1 mV resolution as "Generator Output Voltage, mid range" in the Test Record. Be careful not to alter the output amplitude after the reading is recorded. 18. Set the scope to 200mV/div and note the “Effective Gain” under the ZDxxx tab of the channel setting menu. 19. Divide the reading recorded in step B-17 by Effective Gain noted in step 18. 20.
Instruction Manual Reference Material Specifications Note: Specifications are subject to change without notice. Please refer to the LeCroy website at www.lecroy.com for detailed specification information. Contact LeCroy for Support Use the following regional contacts to find the appropriate support location nearest you.
ZD Series Differential Probes You can also find contact information for our offices on the LeCroy Web site shown for the following regions: Contact Your Local LeCroy Office for Sales and Technical Assistance United States and Canada Phone (Sales, Applications, and Service): 1-800-553-2769 (options 1, 2, and 3, respectively) or 845-425-2000 Fax (Sales, Applications, and Service): 845-578-5985 Email (Sales, Applications, and Service): contact.corp@lecroy.com Web Site: www.lecroy.
Instruction Manual Safety Symbols The following symbols appear on the accessory or in this manual and indicate important safety considerations. Refer to information near this symbol to protect against personal injury or damage to the instrument. The CAUTION sign indicates a potential hazard. It calls attention to a procedure, practice or condition, which, if not followed, could possibly cause damage to the equipment.
ZD Series Differential Probes Safety Requirements CAUTION Avoid personal injury or damage to your accessory or any equipment connected to it by reviewing and complying with the following safety precautions: Use only as intended. The accessory is intended to be used only with the compatible LeCroy instruments. Use of the accessory and/or the equipment it is connected to in a manner other than specified may impair the protection mechanisms. Connect and disconnect properly.
Instruction Manual Compliance Information This LeCroy accessory is CE Compliant and bears the CE mark. CE Declaration of Conformity This ZD Series Probes meet intent of the European Council Directive 2006/95/EEC for Product Safety and 2004/108/EEC for Electromagnetic Compatibility.
ZD Series Differential Probes Probe Input Loading Attaching any probe to a test circuit add some loading to the circuit under test. In most applications the high impedance of the probe, compared to the impedance of the circuit under test, imparts an insignificant load to the test circuit. However, at very high frequencies the capacitive reactance of the probe tip or lead may load the circuit enough to affect the measurement. These probes are designed to minimize these effects at high frequencies.
Instruction Manual Appendix A - Performance Verification Test Record This record can be used to record the results of measurements made during the performance verification of the ZD Series Probes. Photocopy this page and record the results on the copy. File the completed record as required by applicable internal quality procedures. The section in the test record corresponds to the parameters tested in the performance verification procedure.
ZD Series Differential Probes Equipment Used Instrument Model Serial Number Calibration Due Date Oscilloscope Digital Multimeter Function Generator Note: The function generator used in this Performance Verification Procedure is used for making relative measurements. The output of the generator is measured with a DMM or oscilloscope in this procedure. Thus, the generator is not required to be calibrated. Test Record Output Zero Voltage Step A-5.
Instruction Manual Thank you for purchasing a ZD Series Differential Probe.