Care and Maintenance
Cleaning
The exterior of the probe and cable should be cleaned, using a soft cloth moistened with
water. The use of abrasive agents, strong detergents, or other solvents may damage the
probe. Always ensure that the input leads are free of debris.
The probe case is not sealed and should never be immersed in any fluid.
Calibration Interval
The recommended calibration interval is one year. (Performance Verification and Adjustment
Procedures are included in this manual.)
Service Strategy
The ZD Series probe utilizes fine pitch surface mount devices. It is therefore impractical to
attempt to repair in the field. Defective probes must be returned to a LeCroy service facility
for diagnosis and exchange. Defective probes under warranty are repaired or replaced. A
probe that is not under warranty can be exchanged for a factory refurbished probe for a
modest fee. You must return the defective probe in order to receive credit for the probe core.
Returning a Probe for Calibration or Service
Return a probe for calibration or service by contacting your local LeCroy sales representative.
They tell you where to return the product. All returned products should be identified by both
model and serial number. Provide your name and contact number, and a description of the
defect or failure (if possible).