Item Tested
Item Serial Number Date Technician
ZS
Equipment Used
Instrument Model Serial Number Calibration Due
Date
Oscilloscope
Digital Multimeter
Function Generator
Test Record
OUTPUT ZERO VOLTAGE
Step Description Intermediate Data
A-5.
Output Zero (Test limit ≤ ±200 μV)
LF ATTENUATION ACCURACY
Step Description Intermediate Data
B-7 Generator Output Voltage V
B-8 Expected Output Voltage, top range V
B-10 Measured Output Voltage, top range V
B-12
Gain Error, top range (Test Limit ≤ ± 1.0%)
%
B-17 Generator Output Voltage V
B-19 Expected Output Voltage, mid range V
B-21 Measured Output Voltage, mid range V
B-23
Gain Error, top range (Test Limit ≤ ± 1.0%)
%
924282-00 Rev A 25