Manual

TELEDYNE ANALYTICAL INSTRUMENTS
2.0 Operational 2.0 Operational
2.0 Operational 2.0 Operational
2.0 Operational
TheorTheor
TheorTheor
Theor
yy
yy
y
During calibration, the bridge circuit is balanced with the reference
gas; the zero of each range is set with its zero gas and the high end of the
range is adjusted with span gas. The intervening points along the range (or
ranges) of interest will produce a DC electrical signal representative of the
analysis.
2.22.2
2.22.2
2.2
Electronic ComponentsElectronic Components
Electronic ComponentsElectronic Components
Electronic Components
The temperature of the measuring and reference cell block is regulated
to within 0.1°C by means of an electronic proportional controller circuit.
The cell block is mounted on an aluminum heater block and insulated. A
thermistor is used in the circuit to keep the heater block at a predetermined
temperature.
A regulated current supply of approximately 100 milliamperes supplies
stable temperature-independent power to the sensor elements in the cham-
bers. Since the elements are connected to the four arms of the Wheatstone
bridge, a gas with a thermal conductivity different from the reference
creates an imbalance in the bridge in proportion to the thermal conductivity
of the sampled gas. This signal is amplified and displayed.
The output of the final buffer amplifier is normally 1 volt full scale,
and is powered by a standard TAI power supply unit of +15V and -15V.