3 Operation Model 6000A
3-22 Part I
Teledyne Analytical Instruments
–OR –
Go to Ltch and then press either two times or two times.
(Toggle it to N and back to Y.)
3.6 The
Range
Select Function
The
Range
function allows you to manually select the concentration range
of analysis (MANUAL), or to select automatic range switching (AUTO).
In the MANUAL screen, you are further allowed to define the high and low
(concentration) limits of each Range, and select a single, fixed range to run.
CAUTION: If this is a linearized application, the new range must
be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.
3.6.1 Manual (Select/Define Range) Screen
The Manual range-switching mode allows you to select a single, fixed
analysis range. It then allows you to redefine the upper and lower limits, for the
range.
Press
Range
key to start the Range function.
Select range
mode: MANUAL
If above screen displays, use the ∆∇ arrow keys to Select MANUAL, and
press
Enter.
Select range to run
> Ø1 Ø2 Ø3 CAL<
Use the < > keys to select the range: 01, 02, 03, or CAL. Then press
Enter
.
Gas use: SO2
Range: Ø 1 0 %
Use the < > keys to toggle between the Range: low-end field and the
Range: high-end field. Use the
∆∇ keys to change the values of the fields.